首页> 外文期刊>International Journal of Innovative Research in Science, Engineering and Technology >Noninvasive measurement of relative blood oxygen saturation in human retina by using 1μm Fourier Domain Optical Coherence Tomography
【24h】

Noninvasive measurement of relative blood oxygen saturation in human retina by using 1μm Fourier Domain Optical Coherence Tomography

机译:使用1μm傅里叶域光学相干断层扫描技术无创测量人视网膜中的相对血氧饱和度

获取原文
   

获取外文期刊封面封底 >>

       

摘要

The relative blood oxygen saturation measurement has been done in human retina by using Fourier Domain Optical Coherence Tomography (FDOCT) data. A broadband (120 nm bandwidth) light source with central wavelength, 1060 nm was used in FDOCT for deeper penetration in retina. A least squares method was implemented to extract oxy and de-oxy haemoglobin concentrations from the FDOCT data. The measured blood oxygen saturation levels in an artery and veins are comparable with existing methods. Thus, this technique can be used to measure in-vivo relative blood oxygen saturation non-invasive way.
机译:已经通过使用傅立叶域光学相干断层扫描(FDOCT)数据在人体视网膜中进行了相对血氧饱和度测量。 FDOCT使用中心波长为1060 nm的宽带(120 nm带宽)光源在视网膜中进行更深的穿透。实施了最小二乘法从FDOCT数据中提取氧和脱氧血红蛋白浓度。测得的动脉和静脉血氧饱和度水平与现有方法相当。因此,该技术可用于测量体内相对血氧饱和度的非侵入性方式。