首页> 外文期刊>International Journal of Electrochemical Science >An Amperometric Nitrobenzene Electrochemical Sensor Based on Electrochemically Activated Graphite Modified Screen Printed Carbon Electrode
【24h】

An Amperometric Nitrobenzene Electrochemical Sensor Based on Electrochemically Activated Graphite Modified Screen Printed Carbon Electrode

机译:基于电化学活化石墨修饰丝网印刷碳电极的安培硝基苯电化学传感器

获取原文
           

摘要

In the present study involves the fabrication of amperometric sensor for the detection of nitrobenzene(NB) using electrochemically activated graphite (EAG) modified screen printed carbon electrode(SPCE). The EAG modified SPCE was prepared by a simple electrochemical activation of graphite inPBS containing KCl solution at an applied potential of 2.0 V for 300 s. The EAG modified SPCEshowed a good electrocatalytic reduction behavior towards NB with a lower overpotential than that ofother modified SPCEs. Amperometric results reveals that the reduction peak current of NB was linearover the concentrations from 0.3 to 374.5 M. The response time of the sensor was calculated as 5 s. The sensitivity was found as 1.445 AM cm with the detection limit of 0.06 M for NB. Inaddition, the fabricated electrode showed a good selectivity for NB in the presence of nitroaromaticand phenolic compounds with good operational stability.
机译:在本研究中涉及使用电化学活化石墨(EAG)修饰的丝网印刷碳电极(SPCE)制造用于检测硝基苯(NB)的安培传感器。通过在含KCl溶液的PBS中以2.0 V的施加电势将石墨简单电化学活化300 s,即可制备EAG改性的SPCE。 EAG修饰的SPCE对NB表现出良好的电催化还原性能,且过电位低于其他修饰的SPCE。安培测量结果表明,NB的还原峰电流在0.3到374.5 M的浓度范围内呈线性关系。传感器的响应时间为5 s。灵敏度为1.445 AM cm,NB的检出限为0.06M。另外,在硝基芳族化合物和酚类化合物的存在下,所制备的电极对NB具有良好的选择性,且具有良好的操作稳定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号