Determination of Midazolam by Potentiometric PVC membrane and MWCNTs based Carbon Paste Sensors

机译:电位PVC膜和MWCNTs基碳糊传感器测定咪达唑仑。

获取原文
           

摘要

PVC membrane and nano-composite carbon paste potentiometric sensors were made for determinationof midazolam in pharmaceutical formulations. The electrodes respond based on ion-exchangemechanism. The ion-pair used as a sensing element in PVC membrane and carbon paste compositionswas synthesized by interaction of midazolam and tetraphenyl borate. The best PVC membrane sensorresponse was obtained by a membrane composition of 30% PVC, 65% DBP, and 5% ion-pair. Carbonpaste electrode was then designed to have an electrode with the better mechanical resistance. The bestelectrode was composed of 20% ion-pair, 20% paraffin oil, 5% MWCNTs and 55% graphite. Theproposed method was successfully applied in determination of Midazolam in the pharmaceuticalformulations.
机译:制备了PVC膜和纳米复合碳糊电位传感器,用于测定药物制剂中的咪达唑仑。电极基于离子交换机制做出响应。通过咪达唑仑和硼酸四苯酯的相互作用合成了在PVC膜和碳糊组合物中用作传感元件的离子对。最佳的PVC膜传感器响应是通过30%PVC,65%DBP和5%离子对的膜组成获得的。然后将碳糊电极设计为具有更好的机械电阻的电极。最好的电极由20%的离子对,20%的石蜡油,5%的MWCNT和55%的石墨组成。所提出的方法已成功地用于药物制剂中咪达唑仑的测定。