Using ordered mesoporous carbon (OMC) as the sensing material, an amperometric biosensor forascorbic acid (AA) determination was successfully built on a glassy carbon electrode (GCE). Using amicro-electrochemistry system (MECS), the detection capability of the sensor was evaluated. Comparedwith the results obtained for a bare GCE, the AA oxidation overpotential of the OMC detector decreasedby approximately 800 mV. The electro-catalytic oxidation peak currents had a linear relationship withthe AA concentration over the range of 3.5 × 10 ? 7 - 7 × 10 ? 5 mol/L and a detection limit of 1.5 × 10 ? 7mol/L (S/N = 3). Additionally, the sensor was shown to be interference free and exhibited good stabilityand reproducibility.
机译:使用有序介孔碳(OMC)作为传感材料,在玻璃碳电极(GCE)上成功建立了抗坏血酸(AA)测定用的安培生物传感器。使用微电化学系统(MECS),评估了传感器的检测能力。与裸GCE获得的结果相比,OMC检测器的AA氧化超电势降低了约800 mV。电催化氧化峰值电流与AA浓度在3.5×10?范围内呈线性关系。 7-7×10? 5 mol / L,检出限为1.5×10? 7mol / L(S / N = 3)。此外,该传感器被证明是无干扰的,并具有良好的稳定性和可重复性。