This study compares between unmodified carbon paste (CPE; the paste has no ion pair) and polyvinyl chloride (PVC) membrane selective electrodes that were used in potentiometric determination of ketotifen fumarate (KTF), where sodium tetraphenylborate (NaTPB) was used as titrant. The performance characteristics of these sensors were evaluated according to IUPAC recommendations which reveal a fast, stable, and linear response for KTF over the concentration range of 10−7to 10−2 mol L−1. The electrodes show Nernstian slope value of52.51±0.20and51.51±0.25 mV decade−1for CPE and PVC membrane electrodes at 30∘C, respectively. The potential is nearly stable over the pH range 3.0–6.0 and 2.0–7.0 for CPE and PVC membrane electrodes, respectively. Selectivity coefficient values towards different inorganic cations, sugars, and amino acids reflect high selectivity of the prepared electrodes. The electrodes responses at different temperatures were also studied, and long operational lifetime of 12 and 5 weeks for CPE and PVC membrane electrodes, respectively, were found. These are used for determination of ketotifen fumarate using potentiometric titration, calibration, and standard addition methods in pure samples, its pharmaceutical preparations (Zaditen tablets), and biological fluid (urine). The direct potentiometric determination of KTF using the proposed sensors gave recoveries % of98.97±0.53and98.62±0.74with RSD 1.42 and 0.63% for CPE and PVC membrane selective electrodes, respectively. Validation of the method shows suitability of the proposed sensors for use in quality control assessment of KTF. The obtained results were in a good agreement with those obtained using the reported spectrophotometric method.
机译:这项研究比较了未修饰的碳糊(CPE;该糊没有离子对)和聚氯乙烯(PVC)膜选择性电极,后者用于电位测定富马酸酮替芬(KTF),其中四苯硼酸钠(NaTPB)被用作滴定剂。这些传感器的性能特征根据IUPAC的建议进行了评估,该建议揭示了KTF在10-7至10-2 mol L-1的浓度范围内具有快速,稳定和线性的响应。 CPE和PVC膜电极在30∘C时,电极的能斯特斜率分别为52.51±0.20和51.51±0.25 mV decade-1。在CPE和PVC膜电极的pH范围分别为3.0-6.0和2.0-7.0时,电势几乎稳定。对不同的无机阳离子,糖和氨基酸的选择性系数值反映了所制备电极的高选择性。还研究了电极在不同温度下的响应,发现CPE和PVC膜电极的使用寿命分别为12和5周。这些用于通过电位滴定,校准和标准添加方法测定纯样品,其药物制剂(Zaditen片剂)和生物液(尿液)中的富马酸酮替芬。使用建议的传感器直接测定KTF的回收率,CPE和PVC膜选择性电极的回收率分别为98.97±0.53和98.62±0.74,相对标准偏差(RSD)为1.42和0.63%。方法的验证表明所提出的传感器适用于KTF的质量控制评估。获得的结果与使用报道的分光光度法获得的结果吻合良好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号