Accurate knowledge of the beam energy distribution is crucial for particle accelerators,compact medical cyclotrons for the production of radioisotopes in particular. For this purpose,a compact instrument was developed, based on a multi-leaf Faraday cup made of thin aluminum foilsinterleaved with plastic absorbers. The protons stopping in the aluminum foils produce a measurablecurrent that is used to determine the range distribution of the proton beam. On the basis of the protonrange distribution, the beam energy distribution is assessed by means of stopping-power MonteCarlo simulations. In this paper, we report on the design, construction, and testing of this apparatus,as well as on the first measurements performed with the IBA Cyclone 18-MeV medical cyclotron inoperation at the Bern University Hospital.
机译:对束能量分布的准确了解对于粒子加速器,尤其是紧凑型医用回旋加速器,特别是对于放射性同位素的生产至关重要。为了这个目的,开发了一种紧凑的仪器,该仪器基于由薄铝箔制成的多叶法拉第杯和塑料吸收器交错而成。停在铝箔中的质子产生可测量的电流,该电流用于确定质子束的范围分布。根据质子范围分布,通过停止功率蒙特卡洛模拟评估束能量分布。在本文中,我们报告了该设备的设计,构造和测试,以及在伯尔尼大学医院使用IBA Cyclone 18-MeV医用回旋加速器进行的首次测量。