Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters

机译:剂量校准器的预定义校准因子的偏差以及相关的β-γ发射器放射性测量的不准确性

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Aim:To determine whether the predefined calibration factors of the dose calibrators can provide accurate radioactivity measurements of beta-gamma emitters used in routine therapeutic nuclear medicine procedures.Materials and Methods:Two models of dose calibrators were used in the present study for radioactivity measurements of 153Sm ethylenediamine-N, N, N’, N’-tetrakis methylene phosphonic acid (EDTMP) and 177Lu (EDTMP). A known (precalibrated) activity of each of the two beta emitters received by us from our National Supplier for administration to the patients with extensive bony metastases for bone pain palliation, was used for experiments.Results:When we used the manufacturers’ provided dial setting of 450 × 10, each of the dose calibrators underestimated the radioactivity of 177Lu by about 9.0%. Dial settings of 403 × 10 and 408 × 10 for 177Lu on CRC-15R and CRC-ultra dose calibrators respectively were calculated experimentally using an iterative approach. The radioactivity measurements made at these settings provided an excellent agreement with the specified values. Likewise, a dial setting of 230 for each of the two dose calibrators was calculated for 153Sm, which provided a good agreement between the experimentally derived radioactivity values and the certified values. A deviation of ± 5.0% was observed when radioactivity of 177Lu and 153Sm was measured over a wide range (4.0 MBq to 2.1 GBq) for time intervals equivalent to 4.5 half-lives of each of the two radionuclides. A deviation of ± 5% was observed when radioactivity was counted in different dilution volumes and in syringes of varying size.Conclusion:These variations could lead to a cumulative error of about 20.0% toward the inaccuracy in the radioactivity measurements of the beta-gamma emitters and thus predefined calibration factors of the dose calibrators may require experimental re-setting of these parameters and periodic checking to provide accurate radioactivity estimates of beta-gamma emitters in a given clinical setting.
机译:目的:确定剂量校正剂的预定校正因子是否可以提供常规治疗性核医学程序中使用的β-γ发射体的准确放射性测量。材料和方法:本研究中使用了两种剂量校正剂模型来测量放射性153Sm乙二胺-N,N,N',N'-四甲基亚甲基膦酸(EDTMP)和177Lu(EDTMP)。实验使用了我们从国家供应商处接收到的,用于治疗患有广泛性骨转移的骨痛减轻患者的两个β发射体各自的已知(预校准)活性。结果:当我们使用制造商提供的刻度盘设置时在450×10中,每个剂量校正剂低估了177Lu的放射性约9.0%。使用迭代方法,分别通过实验计算出CRC-15R和CRC-超剂量校准器上177Lu的刻度盘设置为403×10和408×10。在这些设置下进行的放射性测量结果与指定值非常吻合。同样,两个剂量校准器的每一个的刻度盘设置为153Sm,计算为230,这在实验得出的放射性值和认证值之间提供了很好的一致性。在宽范围内(4.0 MBq至2.1 GBq)测量177Lu和153Sm的放射性时,观察到±5.0%的偏差,其时间间隔等于两种放射性核素各自的4.5半衰期。当在不同的稀释体积和不同尺寸的注射器中计算放射活性时,观察到的偏差为±5%。结论:这些变异可能导致β-γ发射体放射性测量的不准确度的累积误差约为20.0%。因此,剂量校准器的预定义校准因子可能需要对这些参数进行实验重置,并需要定期检查以在给定的临床环境中提供β-γ发射体的准确放射性估计值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号