Volumetric accuracy of cone-beam computed tomography

机译:锥形束计算机断层扫描的体积精度

获取原文
           

摘要

Purpose This study was performed to investigate the influence of object shape and distance from the center of the image on the volumetric accuracy of cone-beam computed tomography (CBCT) scans, according to different parameters of tube voltage and current. Materials and Methods Four geometric objects (cylinder, cube, pyramid, and hexagon) with predefined dimensions were fabricated. The objects consisted of Teflon-perfluoroalkoxy embedded in a hydrocolloid matrix (Dupli-Coe-Loid TM; GC America Inc., Alsip, IL, USA), encased in an acrylic resin cylinder assembly. An Alphard Vega Dental CT system (Asahi Roentgen Ind. Co., Ltd, Kyoto, Japan) was used to acquire CBCT images. OnDemand 3D (CyberMed Inc., Seoul, Korea) software was used for object segmentation and image analysis. The accuracy was expressed by the volume error (VE). The VE was calculated under 3 different exposure settings. The measured volumes of the objects were compared to the true volumes for statistical analysis. Results The mean VE ranged from ?4.47% to 2.35%. There was no significant relationship between an object's shape and the VE. A significant correlation was found between the distance of the object to the center of the image and the VE. Tube voltage affected the volume measurements and the VE, but tube current did not. Conclusion The evaluated CBCT device provided satisfactory volume measurements. To assess volume measurements, it might be sufficient to use serial scans with a high resolution, but a low dose. This information may provide useful guidance for assessing volume measurements.
机译:目的进行这项研究,以根据不同的管电压和电流参数,研究对象形状和距图像中心的距离对锥束计算机断层扫描(CBCT)扫描的体积精度的影响。材料和方法制造了具有预定尺寸的四个几何对象(圆柱,立方体,金字塔和六边形)。这些物体由嵌入在水胶体基质(Dupli-Coe-Loid TM; GC America Inc.,Alsip,IL,USA)中的特氟隆-全氟烷氧基组成,并包裹在丙烯酸树脂圆柱体组件中。使用Alphard Vega Dental CT系统(Asahi Roentgen Ind。Co.,Ltd,日本京都)来获取CBCT图像。使用OnDemand 3D(Cyber​​Med Inc.,韩国首尔)软件进行对象分割和图像分析。准确度由体积误差(VE)表示。在3种不同的曝光设置下计算VE。将对象的测量体积与真实体积进行比较以进行统计分析。结果平均VE在?4.47%至2.35%之间。物体的形状和VE之间没有显着的关系。发现对象到图像中心的距离与VE之间存在显着的相关性。管电压影响体积测量和VE,但管电流没有影响。结论评估过的CBCT设备可提供令人满意的体积测量结果。要评估体积测量,使用高分辨率但低剂量的串行扫描可能就足够了。该信息可以为评估体积测量提供有用的指导。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号