Intra- and inter-operator reproducibility of automated cloud-based carotid lumen diameter ultrasound measurement

机译:基于云的自动颈动脉内腔直径超声测量的术中和术中可重复性

获取原文
           

摘要

Background Common carotid artery lumen diameter (LD) ultrasound measurement systems are either manual or semi-automated and lack reproducibility and variability studies. This pilot study presents an automated and cloud-based LD measurements software system (AtheroCloud) and evaluates its: (i) intra/inter-operator reproducibility and (ii) intra/inter-observer variability. Methods 100 patients (83 M, mean age: 68 ± 11 years), IRB approved, consisted of L/R CCA artery (200 ultrasound images), acquired using a 7.5-MHz linear transducer. The intra/inter-operator reproducibility was verified using three operator’s readings. Near-wall and far carotid wall borders were manually traced by two observers for intra/inter-observer variability analysis. Results The mean coefficient of correlation (CC) for intra- and inter-operator reproducibility between all the three automated reading pairs were: 0.99 (P 0.0001) and 0.97 (P 0.0001), respectively. The mean CC for intra- and inter-observer variability between both the manual reading pairs were 0.98 (P 0.0001) and 0.98 (P 0.0001), respectively. The Figure-of-Merit between the mean of the three automated readings against the four manuals were 98.32%, 99.50%, 98.94% and 98.49%, respectively. Conclusions The AtheroCloud LD measurement system showed high intra/inter-operator reproducibility hence can be adapted for vascular screening mode or pharmaceutical clinical trial mode.
机译:背景技术颈总动脉内腔直径(LD)超声测量系统是手动或半自动的,缺乏可重复性和变异性研究。这项初步研究提出了一种基于云的自动化LD测量软件系统(AtheroCloud),并对其进行了评估:(i)运营商内部/运营商之间的可重复性以及(ii)运营商内部/运营商之间的可变性。方法100例患者(83 M,平均年龄:68±11岁),经IRB批准,由L / R CCA动脉组成(200幅超声图像),使用7.5 MHz线性换能器获得。使用三个操作员的读数来验证操作员内部/操作员之间的可重复性。由两名观察员手动追踪近壁和远颈壁边界,以进行观察者间/观察者间的变异性分析。结果所有三个自动读取对之间的操作员内和操作员间再现性的平均相关系数(CC)分别为:0.99(P <0.0001)和0.97(P <0.0001)。两个手动阅读对之间的观察者内和观察者间变异的平均CC分别为0.98(P <0.0001)和0.98(P <0.0001)。三个自动读数的平均值与四个手册之间的品质因数分别为98.32%,99.50%,98.94%和98.49%。结论AtheroCloud LD测量系统显示出较高的操作员内/操作员间可重复性,因此可适用于血管筛查模式或药物临床试验模式。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号