首页> 外文期刊>IHJ Cardiovascular Case Reports (CVCR) >Acute dysfunction of tilting disc valve after mitral valve replacement
【24h】

Acute dysfunction of tilting disc valve after mitral valve replacement

机译:更换二尖瓣后碟片阀的急性功能障碍

获取原文
       

摘要

Background Acute prosthetic valve dysfunction is a serious complication following valve replacement surgeries. Extrinsic prosthetic valve obstructions usually limit valve function, thus affecting cardiac output. Materials and methods We report a case of double valve replacements with mono leaflet tilting disc valves and tricuspid repair. Patient had normal prosthetic valve function in immediate post-operative period. She developed episodes of pulseless electrical activity with low cardiac output on the first post-operative day. Patient was re-explored with suspicion of valve dysfunction and pericardial effusion. Transesophageal echocardiogram showed intermittent extrinsic obstruction of mitral prosthesis. Results We observe that acute prosthetic dysfunction can present with unexplained pulse less electrical activity and can be confirmed with transesophageal echocardiography. Conclusion Electro mechanical dissociation and low cardiac output following prosthetic valve replacement may be due to acute prosthetic dysfunction. A high index of suspicion and early intervention can be vital in such patients.
机译:背景技术人工瓣膜置换手术后,急性人工瓣膜功能障碍是严重的并发症。外在的人工瓣膜阻塞通常会限制瓣膜功能,从而影响心输出量。材料和方法我们报告了一例双瓣置换,单瓣斜盘瓣置换和三尖瓣修复。病人在术后立即具有正常的人工瓣膜功能。术后第一天,她出现无脉搏电活动发作,心输出量低。患者再次被怀疑有瓣膜功能不全和心包积液。经食道超声心动图显示二尖瓣假体间断性外在性阻塞。结果我们观察到急性假体功能障碍可表现为原因不明的脉冲,电活动少,经食道超声心动图可证实。结论人工瓣膜置换术后的机械解离和低心输出量可能是由于急性假体功能障碍所致。在此类患者中,高度怀疑和早期干预指数至关重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号