Biomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine

机译:Arcofix系统固定猪脊柱的脊柱生物力学测试

获取原文
       

摘要

Study Design An in vitro biomechanical study. Purpose To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. Overview of Literature Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomechanical properties of these implants have not been sufficiently studied yet. Methods A total of ten porcine specimens (levels Th12-L3) were prepared. The tests were performed for intact, injured, and implanted specimens. In each state, the specimen was subjected to a tension load of a prescribed force, and subsequently, twisted by a given angle. The force load was 200 N. The torsion load had a deformation character, i.e., the control variable was the twisting angle and the measured variable was the moment of a couple. The amplitude of the load alternating cycle was 3°. Another parameter that was evaluated was the area of the hysteresis loop. The area corresponds to the deformation energy which is dissipated during the cycle. Results A statistically significant difference was found between the intact and injured states as well as between the injured and implanted specimens. The statistical evaluation also showed a statistically different value of the hysteresis loop area. In the case of instability, the area decreased to 33% of the physiological value. For the implanted sample, the area increased to 170% of the physiological value. Conclusions The Arcofix implant with its parameters appears to be suitable and sufficiently stable for the treatment of the anterior column of the spine.
机译:研究设计体外生物力学研究。目的评估在用Arcofix植入物固定后完整,受伤和稳定状态下的脊柱节段的机械性能。文献概述几种类型的胸腰椎脊柱损伤需要使用前路器械。 Arcofix板代表了最新一代的角度稳定系统。这些植入物的生物力学特性尚未得到充分研究。方法总共准备了十个猪标本(Th12-L3级)。对完整,受伤和植入的标本进行了测试。在每种状态下,使样品承受规定力的拉力载荷,然后以给定角度扭转。力负载为200N。扭转负载具有变形特性,即,控制变量为扭转角,而测量变量为一对力矩。负载交替循环的幅度为3°。评估的另一个参数是磁滞回线的面积。该面积对应于在循环过程中耗散的变形能。结果在完整状态和受伤状态之间以及受伤和植入的标本之间发现统计学上的显着差异。统计评估还显示,磁滞回线区域的统计值不同。在不稳定的情况下,面积减小到生理值的33%。对于植入的样品,面积增加至生理值的170%。结论Arcofix植入物及其参数对于脊柱前柱的治疗似乎是合适且足够稳定的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号