We demonstrate a new instrument for in situ detection of atmospheric iodineatoms and molecules based on atomic and molecular resonance andoff-resonance ultraviolet fluorescence excited by lamp emission. Theinstrument combines the robustness, light weight, low power consumption andefficient excitation of radio-frequency discharge light sources with thehigh sensitivity of the photon counting technique. Calibration of I2fluorescence is achieved via quantitative detection of the molecule byIncoherent Broad Band Cavity-enhanced Absorption Spectroscopy. Atomic iodinefluorescence signal is calibrated by controlled broad band photolysis ofknown I2 concentrations in the visible spectral range at atmosphericpressure. The instrument has been optimised in laboratory experiments toreach detection limits of 1.2 pptv for I atoms and 13 pptv for I2, forS/N = 1 and 10 min of integration time. The ROFLEX system has beendeployed in a field campaign in northern Spain, representing the firstconcurrent observation of ambient mixing ratios of iodine atoms andmolecules in the 1–350 pptv range.
机译:我们演示了一种新的仪器,用于基于原子和分子共振以及灯发射激发的非共振紫外荧光原位检测大气中的碘原子和分子。该仪器将射频放电光源的鲁棒性,重量轻,功耗低和激发效率高与光子计数技术的高灵敏度结合在一起。 I 2 荧光的校准是通过非相干宽带腔增强吸收光谱法对分子进行定量检测来实现的。通过在大气压下在可见光谱范围内已知I 2 浓度的受控宽带光解来校准原子碘荧光信号。该仪器已经在实验室实验中进行了优化,对于I原子检测限为1.2 pptv,对于I 2 检测限为13 pptv, S / N = 1,积分时间为10分钟。 ROFLEX系统已在西班牙北部的一次野战中部署,这是首次同时观察到碘原子和分子在1–350 pptv范围内的环境混合比。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号