首页> 外文期刊>Annals of laboratory medicine. >Performance of the cobas u 701 Analyzer in Urinary Tract Infection Screening
【24h】

Performance of the cobas u 701 Analyzer in Urinary Tract Infection Screening

机译:cobas u 701分析仪在尿路感染筛查中的性能

获取原文
           

摘要

Background: Negative urine cultures to rule out urinary tract infections (UTI) generate a considerable laboratory workload; thus, a rapid screening test is desirable. We evaluated the performance of a new automated microscopy analyzer, cobas u 701 (Roche Diagnostics International, Rotkreuz, Switzerland) for the screening of UTI, and developed a rule-out strategy to reduce the number of samples requiring culture. We also assessed squamous epithelial cell (SEC) count as a predictor of culture contamination. Methods: In total, 1,604 urine samples from outpatients were analyzed with cobas u 701 and culture. Bacterial (BAC) and white blood cell (WBC) counts were used for sample interpretation. To determine a useful cut-off point to predict negative cultures, we selected the highest sensitivity and specificity values obtained from ROC curves. Diagnostic accuracy by age and gender was evaluated. Results: Urine culture showed growth of ≥10sup4/sup colony forming units (CFU)/mL in 256 samples (16.0%). The highest sensitivity (91.8%) and specificity (68.4%) were obtained for cut-off points of 119 BAC/μL and 22 WBC/μL. The combination of BAC and WBC improved the performance of the rule-out strategy with a low rate of false-negative results (1.5%) and a high negative predictive value (NPV, 97.3%). Fifty-seven percent of the samples would not have required culture. SEC count was a poor predictor of culture contamination. Conclusions: cobas u 701 can substantially reduce the number of urine samples requiring culture, with a low false-negative rate and a high NPV.
机译:背景:排除尿液培养以排除尿路感染(UTI)会产生大量的实验室工作量;因此,需要快速筛选测试。我们评估了新型自动显微分析仪cobas u 701(瑞士罗克勒兹的罗氏诊断国际公司)在UTI筛查中的性能,并制定了一项排除策略以减少需要培养的样品数量。我们还评估了鳞状上皮细胞(SEC)计数作为培养物污染的预测指标。方法:使用cobas u 701和培养物对门诊患者的1604份尿液样本进行了分析。细菌(BAC)和白细胞(WBC)计数用于样品解释。为了确定预测阴性培养的有用的临界点,我们选择了从ROC曲线获得的最高灵敏度和特异性值。评价了按年龄和性别分类的诊断准确性。结果:尿液培养显示256个样品(16.0%)中≥10 4 菌落形成单位(CFU)/ mL的生长。截断点119 BAC /μL和22 WBC /μL的最高灵敏度(91.8%)和特异性(68.4%)。 BAC和WBC的结合提高了排除策略的性能,其假阴性结果率低(1.5%),阴性预测值高(NPV为97.3%)。 57%的样品不需要进行培养。 SEC计数不能很好地预测培养物污染。结论:cobas u 701可以大大减少需要培养的尿液样品的数量,假阴性率低,NPV高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号