首页> 外文期刊>Analytical Sciences >Continuous in vivo Monitoring of Blood Diffusible Calcium Using On-line Microdialysis Sampling Coupled with Flame Atomic Absorption Spectrometry
【24h】

Continuous in vivo Monitoring of Blood Diffusible Calcium Using On-line Microdialysis Sampling Coupled with Flame Atomic Absorption Spectrometry

机译:在线微透析采样结合火焰原子吸收光谱法连续体内监测血液扩散性钙

获取原文
   

获取外文期刊封面封底 >>

       

摘要

A direct, rapid and continuous in vivo monitoring of diffusible calcium in the blood of living rabbits has been developed using microdialysis sampling coupled on-line with flame atomic absorption spectrometry. Microdialysates perfused through implanted microdialysis probes were collected with a sample loop on an injection valve and directly introduced into the flame atomizer by a carrier solution. An ultrapure saline solution (0.9% NaCl, pH 7.2) was used as the perfusion solution at a flow rate of 20 µl min-1 via the microdialysis probe. A 0.1% La solution in 0.5% HNO3 solution was employed as the carrier solution at a nebulizer uptake flow rate of 2.5 ml min-1. The interval for each determination was 2.5 min (2 min of sampling time, 20 s of read time and 10 s of washing time). The performance characteristics of the on-line microdialysis-FAAS system were validated as follows: linearity range, 0 - 100 mg l-1; detection limit (3 σ, n = 7), 3.66 mg l-1; precision (RSD, n = 50), 6.2%. For the evaluation of analytical accuracy, the proposed on-line method was compared with the in vivo no net flux method. The use of an on-line microdialysis-FAAS system permitted the in situ, dynamic and continuous in vivo monitoring of diffusible calcium in the blood of the living rabbits after CaCl2 administration with a temporal resolution of 2.5 min.
机译:使用微透析采样在线结合火焰原子吸收光谱法,已经开发出了对活体兔血液中可扩散钙进行直接,快速和连续的体内监测。通过注入的微透析探针灌注的微透析液通过进样阀上的样品定量环收集,并通过载体溶液直接引入火焰雾化器中。通过微透析探针,将超纯盐溶液(0.9%NaCl,pH 7.2)用作流速为20 µl min-1的灌注溶液。使用在0.5%HNO3溶液中的0.1%La溶液作为载体溶液,雾化器的吸收流速为2.5 ml min-1。每次测定的间隔为2.5分钟(采样时间为2分钟,读取时间为20秒,清洗时间为10秒)。在线微透析-FAAS系统的性能特征验证如下:线性范围为0-100 mg l-1;线性范围为0-100 mg l-1。检出限(3σ,n = 7),3.66 mg l-1;精度(RSD,n = 50),6.2%为了评估分析准确性,将所建议的在线方法与体内无净通量方法进行了比较。在线微透析-FAAS系统的使用允许对CaCl2给药后活体兔血液中的可扩散钙进行原位,动态和连续体内监测,时间分辨率为2.5分钟。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号