摘要

It was found that Cu2+ has a great catalytic effect on p-cresol oxidation by hydrogen peroxide under alkaline conditions, leading to an intense fluorescence signal due to the fast formation of 2,2′-dihydroxy-5,5′-dimethylbiphenyl, an oxidation product of p-cresol. Investigation of the fluorescence spectra of the p-cresol–hydrogen peroxide system demonstrated the catalytic behavior of Cu2+. On this basis, a very simple, sensitive, and selective fluorescent method was established for the determination of trace copper in this study. Under the optimal conditions, the proposed system could respond down to 1.0 × 10?8 mol L?1 of Cu2+ with a linear calibration range of 3.0 × 10?7 mol L?1 to 5.0 × 10?5 mol L?1. The relative standard deviation (RSD) was 2.1% for 7.0 × 10?6 mol L?1Cu2+ (n = 15). The proposed method was successfully applied in the determination of trace Cu2+ in lake, river and swimming pool water samples with satisfactory results. The results obtained by the proposed method are in good agreement with those obtained by a conventional atomic absorption spectroscopy method...
机译:发现Cu2 +在碱性条件下对过氧化氢对对甲酚的氧化具有很大的催化作用,由于2,2'-二羟基-5,5'-二甲基联苯的快速形成,导致强烈的荧光信号对甲酚的产物。对甲酚-过氧化氢体系的荧光光谱研究表明,Cu2 +具有催化作用。在此基础上,本研究建立了一种非常简单,灵敏且选择性的荧光法测定痕量铜。在最佳条件下,所提出的系统可以响应到1.0×10-8 mol L?1的Cu2 +,线性校准范围为3.0×10-7 mol L?1至5.0×10-5 mol L?1。对于7.0×10-6 mol L?1Cu2 +(n = 15),相对标准偏差(RSD)为2.1%。该方法已成功用于湖泊,河流和游泳池水样中痕量Cu2 +的测定,结果令人满意。通过所提出的方法获得的结果与通过常规原子吸收光谱法获得的结果非常吻合...

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号