首页> 外文期刊>American Journal of Analytical Chemistry >Determination of Heavy Metals at Sub-ppb Levels in Water by Graphite Furnace Atomic Absorption Spectrometry Using a Direct Introduction Technique after Preconcentration with an Iminodiacetate Extraction Disk
【24h】

Determination of Heavy Metals at Sub-ppb Levels in Water by Graphite Furnace Atomic Absorption Spectrometry Using a Direct Introduction Technique after Preconcentration with an Iminodiacetate Extraction Disk

机译:亚硝酸双乙酸酯萃取盘预浓缩后,通过直接引入技术通过石墨炉原子吸收光谱法测定水中的亚ppb级重金属

获取原文
       

摘要

A direct analysis method combining an iminodiacetate extraction disk (IED) with graphite furnace atomic absorption spectrometry was developed for the determination of Co, Ni, Cu, Cd, Sn, Pb, and Bi at sub-ppb levels in water. A 100 mL water sample was adjusted to pH 5.6 with nitric acid and a 1 mol?L–1 ammonium acetate solution, and then passed through an IED (diameter, 47 mm; effective filtering diameter, 35 mm) at a flow rate of 80 - 100 mL?min–1 to preconcentrate seven analytes. The IED was dried at 100?C for 20 min in an electric oven, and 110 - 145 small disks, each 2 mm in diameter, were punched out from the IED. A small disk was introduced into the graphite furnace and atomized according to a heating program. For Cd, Sn, Pb, and Bi measurements, Pd was used as a chemical modifier to enhance the absorbances. Calibration was performed using aqueous standard solutions. The detection limits, corresponding to three times the standard deviation (n = 5) of the blank values, were 0.092 μg·L–1 for Co, 0.12 μg·L–1 for Ni, 0.40 μg·L–1 for Cu, 0.077 μg·L–1 for Cd, 0.92 μg·L–1 for Sn, 0.61 μg·L–1 for Pb, and 0.80 μg·L–1 for Bi with an enrichment factor of 140 using a 100-mL water sample. A spike test for the seven analytes in tap water, rainwater, river water, and mineral drinking water showed quantitative recoveries (93% - 108%).
机译:建立了将亚氨基二乙酸酯萃取盘(IED)与石墨炉原子吸收光谱法相结合的直接分析方法,用于测定水中亚ppb含量的Co,Ni,Cu,Cd,Sn,Pb和Bi。用硝酸和1 mol?L–1乙酸铵溶液将100 mL水样品的pH值调节至5.6,然后以80的流速通过IED(直径47 mm;有效过滤直径35 mm)。 -100 mL?min–1以预浓缩7种分析物。将IED在电烤箱中于100°C干燥20分钟,然后从IED上冲出110-145个直径为2 mm的小圆盘。将小圆盘引入石墨炉,并根据加热程序将其雾化。对于Cd,Sn,Pb和Bi的测量,将Pd用作化学改性剂以提高吸光度。使用标准水溶液进行校准。对应于空白值标准偏差(n = 5)的三倍的检出限为:Co 0.092μg·L-1,Ni 0.12μg·L-1,Cu 0.40μg·L-1、0.077使用100 mL水样时,Cd的浓度为μg·L-1,Sn的浓度为0.92μg·L-1,Pb的浓度为0.61μg·L-1,Bi的浓度为140时为0.80μg·L-1。对自来水,雨水,河水和矿泉水中的七种分析物进行加标测试显示定量回收率(93%-108%)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号