首页> 外文期刊>Acta orthopaedica. >Stable fixation of an osseointegated implant system for above-the-knee amputees
【24h】

Stable fixation of an osseointegated implant system for above-the-knee amputees

机译:用于膝上截肢者的骨整合植入系统的稳定固定

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Background and purpose Rehabilitation of patients with transfemoral amputations is particularly difficult due to problems in using standard socket prostheses. We wanted to assess long-term fixation of the osseointegrated implant system (OPRA) using radiostereometric analysis (RSA) and periprosthetic bone remodeling. Methods 51 patients with transfemoral amputations (55 implants) were enrolled in an RSA study. RSA and plain radiographs were scheduled at 6 months and at 1, 2, 5, 7, and 10 years after surgery. RSA films were analyzed using UmRSA software. Plain radiographs were graded for bone resorption, cancellization, cortical thinning, and trabecular streaming or buttressing in specifically defined zones around the implant. Results At 5 years, the median (SE) migration of the implant was –0.02 (0.06) mm distally. The rotational movement was 0.42 (0.32) degrees around the longitudinal axis. There was no statistically significant difference in median rotation or migration at any follow-up time. Cancellizat...
机译:背景和目的由于使用标准的套筒假体存在问题,经股截肢患者的康复特别困难。我们希望使用放射立体分析(RSA)和假体周围骨改建来评估骨整合植入物系统(OPRA)的长期固定。方法对51例经股截肢的患者(55个植入物)进行了RSA研究。 RSA和X线平片定于术后6个月以及术后1、2、5、7和10年。使用UmRSA软件分析RSA胶片。对普通X射线照片进行分级,以便在植入物周围的特定区域内进行骨吸收,消融,皮质变薄以及小梁流或支撑。结果5年时,植入物向远端的中位(SE)迁移为–0.02(0.06)mm。绕纵轴的旋转运动为0.42(0.32)度。在任何随访时间,中位旋转或迁移均无统计学差异。取消...

著录项