2D measurements of cup orientation are less reliable than 3D measurements

机译:杯子方向的2D测量不如3D测量可靠

获取原文
           

摘要

Background and purpose — 2D analysis of metal-on-metal (MoM) hip arthroplasty (HA) has been conducted in several large series on conventional radiographs with the use of Ein Bild Roentegen Analyse (EBRA) software, but there have been no comparisons with 3D analysis in the literature. The main aim of this study was to quantify the agreement in measurements of cup version of large-diameter MoM hips obtained by EBRA and by 3D computed tomography (3D-CT). The secondary aim was to quantify the agreement for cup inclination. Lastly, we wanted to determine the inter- and intra-observer reliability of both methods.Patients and methods — 87 MoM hips in 81 patients were analyzed for cup inclination and version in 2D on conventional radiographs using EBRA software. The results were compared with 3D measurements using CT.Results — Cup version was underestimated by EBRA when compared to 3D-CT, by 6° on average with the pelvis supine and by 8° on average with the pelvis orientated to the anterior pelvic plane (APP). Fo...
机译:背景与目的—使用Ein Bild Roentegen Analyze(EBRA)软件在常规X射线照片上对金属对金属(MoM)髋关节置换术(HA)进行了多个大型系列的二维分析,但没有与之进行比较。文献中的3D分析。这项研究的主要目的是量化通过EBRA和3D计算机断层扫描(3D-CT)获得的大直径MoM髋关节杯形测量的一致性。第二个目的是量化杯子倾斜度的一致性。最后,我们想确定两种方法在观察者之间和观察者之间的可靠性。患者和方法-使用EBRA软件在常规X射线照片上分析了81例患者的87个MoM髋关节的杯倾斜度和二维版本。结果与使用CT进行的3D测量结果进行了比较。结果-与3D-CT相比,EBRA低估了杯型,骨盆仰卧位平均低6°,而骨盆面向前骨盆平面平均低8°( APP)。佛...