首页> 外文期刊>Acoustical science and technology >Blood flow noise transducer for detecting intracranial vascular deformations
【24h】

Blood flow noise transducer for detecting intracranial vascular deformations

机译:用于检测颅内血管变形的血流噪声传感器

获取原文
           

摘要

References(15) A transducer using PVDF film, which is placed on the eyelid, has been designed to measure intracranial blood flow noise, called “bruit”, as a screening test for detecting intracranial vascular deformations such as aneurysms, stenosis, and malformations. First, the mechanical impedance of the eye was measured. Then, the transducer performance on the eye was evaluated in terms of sensitivity versus frequency while varying the parameters of the transducer components. The prototype transducer fabricated had a sensitivity of −76dB(0dB=1V/(dyn/cm2)) at frequencies of up to around 1, 000Hz. It was successfully used in a neurosurgical clinic to measure the bruit signals, and was proved to be effective for detecting cerebral aneurysms and stenosis.
机译:参考文献(15)设计了一种使用PVDF膜的传感器,该传感器放置在眼睑上,用于测量颅内血流噪声(称为“ bruit”),作为用于检测颅内血管变形(如动脉瘤,狭窄和畸形)的筛查测试。首先,测量眼睛的机械阻抗。然后,在改变传感器组件参数的同时,根据灵敏度与频率的关系评估了传感器在眼睛上的性能。所制造的原型换能器在高达1,000Hz的频率下具有−76dB(0dB = 1V /(dyn / cm2))的灵敏度。它已成功地在神经外科诊所中用于测量残渣信号,并被证明对检测脑动脉瘤和狭窄有效。