Minimally spaced electrode positions for multi-functional chest sensors: ECG and respiratory signal estimation

机译:多功能胸部传感器的最小间距电极位置:ECG和呼吸信号估计

获取原文
           

摘要

Unobtrusive medical instrumentation is a key in continuous patient monitoring. To increase compliance, multi-functional sensor concepts and measurement sites different from gold-standards are used. In this work, we aim to combine both approaches. We focus on minimally spaced electrode positions with high signal correlations to gold-standards. We present twofold experimental data from six and eleven healthy volunteers and provide chest positions with individual correlations up to 0.83 ± 0.06 for ECG and 0.73 ± 0.28 for the respiratory frequency. Using a performance index, we assess positions with correlations up to 0.77 ± 0.12 for ECG and 0.65 ± 0.35 for the respiratory frequency with 24 mm electrode distance.
机译:稳健的医疗仪器是持续监测患者的关键。为了提高合规性,使用了不同于金标准的多功能传感器概念和测量位置。在这项工作中,我们旨在将两种方法结合起来。我们专注于与金标准具有高信号相关性的最小间距的电极位置。我们提供了来自六名和十一名健康志愿者的双重实验数据,并提供了与心电图的个人相关性高达0.83±0.06,对呼吸频率的个人相关性高达0.73±0.28的胸部位置。使用性能指数,我们评估电极位置为24 mm时与ECG的相关性高达0.77±0.12,对于呼吸频率的相关性高达0.65±0.35。