Cadaveric analysis of an accelerometer-based portable navigation device for distal femoral cutting block alignment in total knee arthroplasty

机译:基于加速度计的便携式导航设备在全膝关节置换术中股骨远端切割块对齐的尸体分析

获取原文
           

摘要

Femoral intramedullary guides have been shown to be insufficiently accurate in creating a distal femoral resection perpendicular to the mechanical axis in total knee arthroplasty (TKA), as they make assumptions regarding the difference between the patient's femoral mechanical and anatomical angles. The aim of this cadaveric study was to validate the accuracy of a portable accelerometer-based navigation device for alignment of the distal femoral cutting block in TKA. Twenty-nine trials were performed on five cadaveric specimens (hip-to-ankle), in which the distal femoral cutting block was placed using the KneeAlign 2? navigation device. For each specimen, a preoperative “target” was assigned for varus/valgus and flexion/extension alignment of the cutting block. The actual alignment of each cutting block was then measured using the ORTHOsoft Computer Assisted Surgery (CAS) system. The mean absolute difference between the preoperative target and the alignment of the cutting block was 0.83?±?0.60° for varus/valgus, and 0.83?±?0.83° for flexion/extension. The KneeAlign 2? navigation device can set and align the distal femoral resection guide with the same accuracy as a large-console CAS system, thus demonstrating that portable accelerometer-based navigation can be used reliably in total knee arthroplasty.
机译:在全膝关节置换术(TKA)中,股骨髓内导管在创建垂直于机械轴的股骨远端切除术时表现得不够准确,因为它们假设患者的股骨力学角度与解剖角度之间存在差异。这项尸体研究的目的是验证基于便携式加速度计的导航设备在TKA中对齐股骨远端切割块的准确性。在五个尸体标本(髋到踝)上进行了29项试验,其中使用KneeAlign 2™放置了股骨远端切块。导航设备。对于每个标本,为切割块的内翻/外翻和屈曲/伸展对准指定术前“目标”。然后使用ORTHOsoft计算机辅助手术(CAS)系统测量每个切割块的实际对齐情况。术前目标与切割块对齐的平均绝对差内翻/外翻为0.83?±?0.60°,屈曲/伸展为0.83?±?0.83°。 KneeAlign 2?导航设备可以以与大控制台CAS系统相同的精度设置和对齐股骨远端切除术,因此证明了基于便携式加速度计的导航可以可靠地用于全膝关节置换术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号