Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma

机译:日本原发性开角型青光眼患者使用Goldmann压平眼压计重复测量眼压的波动相关因素

获取原文
       

摘要

Purpose: The aim of this study was to determine whether fluctuations in intraocular pressure (IOP) occur as a result of the order of IOP measurements or successive IOP measurements in patients with glaucoma and, if so, identify the factors causing these fluctuations. Patients and methods: Four hundred twenty-eight eyes of 214 Japanese patients with primary open-angle glaucoma (POAG) were enrolled. Patients treated with beta-blockers or prostaglandin analogs alone were included. Additionally, in the IOP measurements by noncontact tonometer, the same cases of IOP of the right and left eyes prior to this study were included in this study. Four successive IOP measurements were carried out using a Goldmann applanation tonometer as follows: IOP was measured in the first eye (right or left) and then in the fellow eye and IOP was again measured in the first eye and then in the fellow eye. Repeated-measures analysis of variance was used to test the differences in IOP between successive measurements. Generalized linear mixed models were used to test differences in IOP measurements between the right and the left eyes on repeated applanation tonometry and according to the order of measurement. Conditional binomial logistic regression analysis was used to identify factors associated with fluctuating repeated applanation tonometry measurements. A P -value of 0.05 was considered statistically significant. Results: IOP values decreased significantly according to the number of measurements (13.8–13.0; P 0.001–0.036, respectively). There was no significant difference in IOP measurements between the right and left eyes. The first IOP measurement was significantly higher than the fourth measurement ( P =0.038); however, there was no significant difference between other combinations. The use of a prostaglandin analog was the only significant contributor to fluctuating IOP measurements ( P =0.002). Conclusion: IOP measured in the first eye, either right or left, was higher than that measured in the fellow eye in Japanese patients with POAG. The use of a prostaglandin analog may be associated with fluctuating IOP on repeated applanation tonometry.
机译:目的:本研究的目的是确定青光眼患者眼内压(IOP)的波动是否是由于眼压测量顺序或连续眼压测量的结果而发生的,如果是,则确定引起这些波动的因素。患者和方法:纳入214例日本原发性开角型青光眼(POAG)患者的248眼。包括仅接受β受体阻滞剂或前列腺素类似物治疗的患者。此外,在通过非接触式眼压计进行的IOP测量中,本研究之前包括右眼和左眼的IOP的相同病例。使用Goldmann扁平眼压计按以下步骤进行四次连续的IOP测量:先在第一只眼(右或左)中测量IOP,然后在另一只眼中测量IOP,再次在第一只眼中然后在另一只眼中测量IOP。重复测量方差分析用于测试连续测量之间的IOP差异。广义线性混合模型用于根据重复压平眼压计并根据测量顺序来测试右眼和左眼之间的IOP测量差异。条件二项式logistic回归分析用于确定与反复压平眼压测量值波动相关的因素。 P值<0.05被认为具有统计学意义。结果:根据测量次数,IOP值显着下降(分别为13.8-13.0; P <0.001-0.036)。右眼和左眼之间的IOP测量没有显着差异。第一次IOP测量显着高于第四次测量(P = 0.038);但是,其他组合之间没有显着差异。前列腺素类似物的使用是引起眼压波动的唯一重要因素(P = 0.002)。结论:日本POAG患者的第一只眼(右眼或左眼)的IOP高于另一只眼的IOP。前列腺素类似物的使用可能与反复压平眼压计上的眼压波动有关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号