Intraocular pressure measurements with Goldmann applanation tonometry and dynamic contour tonometry in eyes after IntraLASIK or LASEK

机译:在进行IntraLASIK或LASEK手术后使用Goldmann压平眼压计和动态轮廓眼压计测量眼内压

获取原文
           

摘要

Background and methods: Myopic photorefractive surgery induces a reduction in central corneal thickness, which may lead to underestimation of intraocular pressure. This retrospective clinical study compared intraocular pressure measurements obtained by Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT-Pascal) in eyes undergoing myopic intralaser-assisted in situ keratomileusis (IntraLASIK) or laser-assisted subepithelial keratomileusis (LASEK).Results: Of a total of 51 eyes, 21 underwent LASEK and 30 underwent IntraLASIK. By GAT, mean preoperative intraocular pressure was 16.2 ± 1.99 mmHg and postoperatively was 10.84 ± 1.45 mmHg. By DCT, mean preoperative intraocular pressure was 15.9 ± 2.08 mmHg and postoperatively was 16.1 ± 2.3 mmHg. Both preoperative and postoperative differences between measurements made by GAT and DCT were found to be statistically significant (P < 0.04 and P < 0.01, respectively). GAT and DCT readings were unaffected by type of surgery (P = 0.74 and P = 0.46, respectively).Conclusion: Postoperative GAT measurements were lower than those obtained by DCT. The difference between preoperative and postoperative DCT measurements was minimal, so DCT may be preferable for the measurement of intraocular pressure in eyes undergoing myopic IntraLASIK or LASEK.
机译:背景和方法:近视屈光手术会导致中央角膜厚度的减少,这可能导致低估眼内压。这项回顾性临床研究比较了Goldmann压平眼压计(GAT)和动态轮廓眼压计(DCT-Pascal)在近视激光辅助原位角膜磨镶术(IntraLASIK)或激光辅助上皮下角膜磨镶术(LASEK)眼中的眼压测量结果。在总共51眼中,有21眼接受了LASEK,而30眼接受了IntraLASIK。术前平均眼压为16.2±1.99 mmHg,术后平均眼压为10.84±1.45 mmHg。通过DCT,术前平均眼压为15.9±2.08 mmHg,术后平均眼压为16.1±2.3 mmHg。发现GAT和DCT测量值之间的术前和术后差异在统计学上均具有统计学意义(分别为P <0.04和P <0.01)。 GAT和DCT读数不受手术类型的影响(分别为P = 0.74和P = 0.46)。结论:术后GAT测量值低于DCT测量值。术前和术后DCT测量之间的差异很小,因此DCT可能更适合于测量近视IntraLASIK或LASEK眼睛的眼内压。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号