首页> 外文期刊>Canadian Urological Association Journal >Delayed presentation of a detached resectoscope beak and treatment with thulium laser
【24h】

Delayed presentation of a detached resectoscope beak and treatment with thulium laser

机译:延迟出现分离式电切镜喙和and激光治疗

获取原文
           

摘要

Intraoperative detachment of a resectoscope beak is an extremely rare and unusual event. Retrieval of the ceramic beak can be difficult due to the sheath’s large size relative to the urethral lumen. Our report describes the retrieval of a calcified ceramic resectoscope beak from the urinary bladder 4 years after transurethral resection of the prostate (TURP). The calcified beak was successfully fragmented into 3 separate pieces with a thulium laser. This method allowed for safe removal of each fragment through the urethra.
机译:术中切除直肠镜喙是非常罕见且不寻常的事件。由于鞘管相对于尿道腔较大,因此很难取回陶瓷喙。我们的报告描述了经尿道前列腺电切术(TURP)4年后从膀胱中取出的钙化陶瓷电切镜喙。用th激光将钙化的喙成功地分成3个独立的部分。这种方法可以安全地通过尿道清除每个碎片。