Connecting accuracy of interchanged prosthetic abutments to different dental implants using scanning electron microscopy

机译:使用扫描电子显微镜将互换后的义齿基台连接到不同牙种植体的连接精度

获取原文
           

摘要

The present study evaluated the interchangeability of prosthetic components for external hexagon implants by measuring the precision of the implant/abutment (I/A) interface with scanning electron microscopy. Ten implants for each of three brands (SIN, Conex?o, Neodent) were tested with their respective abutments (milled CoCr collar rotational and non-rotational) and another of an alternative manufacturer (Microplant) in randomly arranged I/A combinations. The degree of interchangeability between the various brands of components was defined using the original abutment interface gap with its respective implant as the benchmark dimension. Accordingly, when the result for a given component placed on an implant was equal to or smaller then that gap measured when the original component of the same brand as the implant was positioned, interchangeability was considered valid. Data were compared with the Kruskal-Wallis test at 5% significance level. Some degree of misfit was observed in all specimens. Generally, the non-rotational component was more accurate than its rotational counterpart. The latter samples ranged from 0.6-16.9 μm, with a 4.6 μm median; and the former from 0.3-12.9 μm, with a 3.4 μm median. Specimens with the abutment and fixture from Conex?o had larger microgap than the original set for SIN and Neodent (p
机译:本研究通过用扫描电子显微镜测量植入物/基台(I / A)界面的精度,评估了外部六角形植入物的假体组件的互换性。对三个品牌(SIN,Conex?o,Neodent)中的每个品牌的十个植入物分别使用各自的基台(铣削的CoCr颈圈旋转和不旋转)进行了测试,另外另一家制造商(Microplant)以随机排列的I / A组合进行了测试。使用原始的基台界面间隙及其相应的植入物作为基准尺寸,定义了各个品牌组件之间的互换性程度。因此,当放置在植入物上的给定组件的结果等于或小于在放置与植入物相同品牌的原始组件时测量的间隙时,可互换性被认为是有效的。将数据与5%显着性水平的Kruskal-Wallis检验进行比较。在所有标本中都观察到一定程度的失配。通常,非旋转组件比其旋转组件更准确。后面的样本范围为0.6-16.9μm,中位数为4.6μm;前者为0.3-12.9μm,中位数为3.4μm。来自Conex?o的带有基台和固定装置的标本具有比SIN和Neodent的原始设置更大的微间隙(p

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号