Comparison of three hand dynamometers in relation to the accuracy and precision of the measurements

机译:三种测功机在测量精度和精确度方面的比较

获取原文
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: Given the variety of available hand-held dynamometers and their different handle shapes, reliability studies are needed. Objectives: To compare the accuracy and reliability between three different hand-held dynamometers and analyze the influence of their handles on grip strength. Methods: The tests were performed with the Jamar? dynamometer, the Takei? dynamometer and the EMG System Manual Transducer with modified handle. Eighteen healthy volunteers aged 20.0±1.3 years without history of musculoskeletal disorders or trauma in the evaluated limbs were included. Data normality was tested using the Shapiro-Wilk test. To verify possible differences between the dynamometers, repeated measures ANOVA was administered, followed by Tukey post-hoc tests. Reliability between measurements was evaluated using intraclass correlation coefficient (ICC) and agreement was tested using Bland and Altman plots. The dynamometers calibration process was evaluated using linear regressions. RESULTS: We observed statistically significant differences on the female group between the Jamar? and the Takei? dynamometers (females p
机译:背景:鉴于各种可用的手持式测功机及其手柄形状不同,需要进行可靠性研究。目的:比较三种不同的手持测功机的准确性和可靠性,并分析其手柄对握力的影响。方法:使用Jamar?进行测试。测力计,武井?测功机和带有改良手柄的EMG系统手动传感器。纳入了18位20.0±1.3岁的健康志愿者,这些志愿者在被评估的四肢中没有肌肉骨骼疾病或外伤史。数据正态性使用Shapiro-Wilk检验进行了检验。为了验证测功机之间可能存在的差异,重复测量了ANOVA,然后进行了Tukey事后测试。使用类内相关系数(ICC)评估测量之间的可靠性,并使用Bland和Altman图测试一致性。使用线性回归评估测功机的校准过程。结果:我们观察到在Jamar?还有武井测力计(女p

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号