In Vivo Characterization of Titanium Implants Coated with Synthetic Hydroxyapatite by Electrophoresis

机译:合成羟基磷灰石涂层钛植入物的体内表征

获取原文
           

摘要

This study compared in vivo the performances of commercially pure titanium (cp Ti) screw dental implants either uncoated or coated with synthetic hydroxyapatite (HA) by electrophoresis. The HA coating was characterized by scanning electron microscopy, energy dispersive spectroscopy (EDS) and Fourier-transform infrared (FT-IR) spectroscopy. Well-adhered carbonated-hydroxyapatite layers (4- to-8-μm-thick) were obtained. In vivo tests were carried out by insertion of both uncoated and HA-coated implants into rabbit tibiae for 8 or 12 weeks. Histomorphometric analysis was performed by scanning electron microscopy with the aid of image-processing software. Results showed significantly greater bone-implant contact for HA-coated implants (p0.05) between both types of implants, although cp Ti had numerically higher percentage of bone content than HA-coated implants. In conclusion, the HA-coated implants had better performance regarding the bone-implant contact area than the uncoated implants; coating by electrophoresis proved to be a valuable process to coat metallic implants with an osteoconductive material such as hydroxyapatite.
机译:这项研究在体内通过电泳比较了未涂覆或未涂覆合成羟基磷灰石(HA)的商业纯钛(cp Ti)螺钉式牙科植入物的性能。 HA涂层通过扫描电子显微镜,能量色散光谱(EDS)和傅立叶变换红外光谱(FT-IR)光谱进行表征。获得了粘附良好的碳酸羟基磷灰石层(厚度为4至8μm)。通过将未包被的和HA包被的植入物插入兔子胫骨中进行8或12周来进行体内测试。借助图像处理软件通过扫描电子显微镜进行组织形态计量学分析。结果表明,两种类型的植入物之间的HA涂层植入物的骨-植入物接触量均显着增加(p0.05),尽管cp Ti的骨含量百分比数值高于HA涂层植入物。总之,与未涂层的种植体相比,HA涂层的种植体在骨-种植体接触面积方面具有更好的性能。电泳涂覆被证明是用骨导电材料(例如羟基磷灰石)涂覆金属植入物的有价值的过程。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号