首页> 外文期刊>Brazilian Dental Journal >An in vivo radiographic evaluation of the accuracy of Apex and iPex electronic Apex locators
【24h】

An in vivo radiographic evaluation of the accuracy of Apex and iPex electronic Apex locators

机译:对Apex和iPex电子Apex定位器的准确性进行体内射线照相评估

获取原文
           

摘要

The aim of this study was to evaluate in vivo the clinical applicability of two electronic apex locators (EALs) - Apex (Septodont) and iPex (NSK) - in different groups of human teeth by using radiography. The working lengths (WLs) of 100 root canals were determined electronically. The EAL to be used first was chosen randomly and a K-file was inserted into the root canal until the EAL display indicated the location of the apical constriction (0 mm). The K-file was fixed to the tooth and a periapical radiograph was taken using a radiographic film holder. The K-file was removed and the WL was measured. The same procedure was repeated using the other EAL. Radiographs were examined with the aid of a light-box with lens of ×4 magnification by two blinded experienced endodontists. The distance between the file tip and the root apex was recorded as follows: (A) +1 to 0 mm, (B) -0.1 to 0.5 mm, (C) -0.6 to 1 mm, (D) -1.1 to 1.5 mm, and (E) -1.6 mm or greater. For statistical purposes, these scores were divided into 2 subgroups according to the radiographic apex: acceptable (B, C, and D) and non-acceptable (A and E). Statistically significant differences were not found between the results of Apex and iPex in terms of acceptable and non-acceptable measurements (p>0.05) or in terms of the distance recorded from file tip and the radiographic apex (p>0.05). Apex and iPex EALs provided reliable measurements for WL determination for endodontic therapy.
机译:这项研究的目的是通过放射线照相法评估两种电子先头定位器(EAL)-Apex(Septodont)和iPex(NSK)在体内的临床适用性。用电子方法确定100个根管的工作长度(WLs)。随机选择首先要使用的EAL,然后将K形锉插入根管,直到EAL显示屏显示根尖收缩的位置(0毫米)。将K形锉固定在牙齿上,并用X射线胶片保持器拍摄根尖周X光片。删除K文件并测量WL。使用其他EAL重复相同的步骤。两名盲人经验丰富的牙髓医师借助带有×4倍透镜的灯箱检查了射线照片。锉尖与根尖之间的距离记录如下:(A)+1至0毫米,(B)-0.1至0.5毫米,(C)-0.6至1毫米,(D)-1.1至1.5毫米,并且(E)-1.6 mm或更大。出于统计目的,根据射线照相顶点将这些分数分为2个子组:可接受(B,C和D)和不可接受(A和E)。在Apex和iPex的结果之间,在可接受和不可接受的测量值(p> 0.05)或从笔尖到X射线影像顶点的记录距离(p> 0.05)之间均未发现统计学上的显着差异。 Apex和iPex EAL为牙髓治疗的WL确定提供了可靠的测量方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号