The comparison of thermal tissue injuries caused by ultrasonic scalpel and electrocautery use in rabbit tongue tissue.

机译:超声刀和电烙术对兔舌组织造成的热组织损伤的比较。

获取原文
           

摘要

The aim of this study compares to the increase in tissue temperature and the thermal histological effects of ultrasonic scalpel, bipolar and unipolar electrosurgery incisions in the tongue tissue of rabbits. This study evaluates the histopathological changes related to thermal change and the maximum temperature values in the peripheral tissue brought about by the incisions carried out by the three methods in a comparative way.To assess thermal tissue damage induced by the three instruments, maximum tissue temperatures were measured during the surgical procedure and tongue tissue samples were examined histopathologically following the surgery. The mean maximum temperature values of the groups were 93.93±2.76 Co for the unipolar electrocautery group, whereas 85.07±5.95 Co for the bipolar electrocautery group, and 108.23±7.64 Co for the ultrasonic scalpel group.There was a statistically significant relationship between the increase in maximum temperature values and the separation among tissue layers, edema, congestion, necrosis, hemorrhage, destruction in blood vessel walls and fibrin accumulation, and between the existence of fibrin thrombus and tissue damage depth (p
机译:本研究的目的是比较兔舌组织中组织温度的升高以及超声刀,双极和单极电外科切口的热组织学效应。这项研究以比较的方式评估了由三种方法进行的切口引起的与热变化和周围组织最高温度相关的组织病理学变化。为了评估这三种仪器引起的热组织损伤,最高组织温度为在手术过程中测量舌头组织样本,并在手术后组织学检查舌组织样本。单极电烙组各组的平均最高温度值分别为93.93±2.76 Co,双极电烙组为85.07±5.95 Co,超声手术刀组为108.23±7.64 Co.最大温度值和组织层之间的分离,水肿,充血,坏死,出血,血管壁破坏和纤维蛋白积聚,以及纤维蛋白血栓的存在与组织损伤深度之间的关系

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号