Abstract Objective: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. Methods: The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves. Results: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition. Conclusion: Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained.
机译:摘要目的:根据生理学文献综述的体内研究,对有关心室变化与左心室收缩和舒张持续时间的ISO 5840标准进行补充。方法:将三个体内研究的收缩期和舒张期与ISO 5840:2010和ISO 5840-2:2015提出的用于人工心脏瓣膜水动力性能评估的收缩期进行比较。结果:根据所分析的体内研究,ISO 5840标准提出的收缩期似乎对于每分钟45次和120次搏动(bpm)是一致的,并且在70 bpm的情况下显示出不同的结果。结论:获得了关于不同心率的左心室模型手术的实际验证的信息。