首页> 外文期刊>BMC Anesthesiology >Assessment of agreement and interchangeability between the TEG5000 and TEG6S thromboelastography haemostasis analysers: a prospective validation study
【24h】

Assessment of agreement and interchangeability between the TEG5000 and TEG6S thromboelastography haemostasis analysers: a prospective validation study

机译:TEG5000和TEG6S血栓弹性成像止血分析仪之间一致性和互换性的评估:一项前瞻性验证研究

获取原文
   

获取外文期刊封面封底 >>

       

摘要

TEG6S? and TEG5000? (Haemonetics Corp, USA) are haemostasis analysers that measure viscoelasticity properties of whole blood. Both use different mechanisms to assess similar components of the coagulation process. The aim of this study was to assess agreement and interchangeability between the TEG6S and TEG5000 analysers. 3.5?mL whole blood was collected from 25 adult patients in a tertiary intensive care unit (ICU). Analysis was performed using TEG6S and TEG5000 haemostatic platforms. Agreement between platforms was measured using Lin’s concordance coefficient (Lin’s CC), further validated using intraclass correlation coefficients and reduced major axis regression (RMAR). Sixteen (64%) patients were male; mean (range) age: 59yo (23–86). TEG6S and TEG5000 systems were broadly interchangeable. The majority of TEG variables demonstrated almost perfect or substantial agreement and minimal proportional bias (maximum amplitude demonstrated a fixed bias). LY30%, however, demonstrated poor agreement and a proportional bias. Lin’s CC coefficients (95% CI, RMAR slope, intercept) between TEG6S and TEG5000 variables were: R time: 0.78 (0.64–0.92, 0.76, 0.92); K time: 0.82 (0.69–0.94, 1.30, ??0.93); alpha angle: 0.79 (0.64–0.95, 1.04, ??1.43); maximum amplitude (MA): 0.90 (0.83–0.96, 0.99, ??5.0); LY30%: 0.34 (0.1–0.58, 0.43, 0.04). Adult patients with critical illness demonstrate almost perfect agreement in the R time and MA, substantial agreement in K time and alpha angle, but poor agreement in LY30%, as measured by the TEG6S and TEG5000 analysers. With the exception of LY30%, the TEG6S and TEG5000 platforms appear interchangeable. This has important implications for use in clinical practice and multi-site research programs. ANZCRT number: 12617000062325 , registered 12/Jan17. Retrospectively registered.
机译:TEG6S?和TEG5000? (Haemonetics Corp,美国)是用于测量全血粘弹性特性的止血分析仪。两者都使用不同的机制来评估凝血过程的相似组成部分。这项研究的目的是评估TEG6S和TEG5000分析仪之间的一致性和互换性。在三级重症监护病房(ICU)中从25名成年患者中收集了3.5?mL全血。使用TEG6S和TEG5000止血平台进行分析。平台之间的一致性使用Lin的一致性系数(Lin's CC)进行了测量,并使用类内相关系数和减少的主轴回归(RMAR)进行了进一步验证。 16名(64%)患者为男性;平均年龄:59岁(23–86)。 TEG6S和TEG5000系统可广泛互换。大多数TEG变量显示几乎完美或基本一致,并且比例偏差最小(最大振幅表明固定偏差)。 LY30%的人表现出较差的一致性和成比例的偏见。在TEG6S和TEG5000变量之间,林的CC系数(95%CI,RMAR斜率,截距)为:R时间:0.78(0.64-0.92,0.76,0.92); K时间:0.82(0.69-0.94,1.30,≤0.93);阿尔法角:0.79(0.64-0.95,1.04,?? 1.43);最大振幅(MA):0.90(0.83-0.96,0.99,?? 5.0); LY30%:0.34(0.1-0.58,0.43,0.04)。根据TEG6S和TEG5000分析仪的测量,成年重症患者在R时间和MA上表现出近乎完美的一致性,在K时间和α角上表现出明显的一致性,但在LY30%中表现出差的一致性。除LY30%外,TEG6S和TEG5000平台似乎可以互换。这对于在临床实践和多站点研究计划中使用具有重要意义。 ANZCRT编号:12617000062325,17年1月12日注册。追溯注册。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号