ObjectiveStethoscope is a medical device universally used by health care workers. Stethoscope may transmit pathogens among patients and health care workers if it is not disinfected. The objective of this study was to, determine the level of stethoscope contamination used by health care workers, survey the practices of disinfecting the stethoscope, identify various microorganisms and assess their role as potential pathogens and determine the effectiveness of 70% ethanol as a disinfecting agent. ResultsThis was a cross-sectional study conducted in the department of Microbiology, Chitwan Medical College, Bharatpur, Nepal. Stethoscopes of 122 health care workers from different departments were included in this study. Out of a total 122 diaphragms, 88 (72.1%) were colonized. Only 71 (58.1%) bells and 152 earpieces (66.2%) were contaminated. Micrococcus and coagulase negative staphylococci were predominantly isolated species. The contamination was lowest among stethoscopes cleaned after touching every patient (11.5%) and the difference is statistically significant (P
机译:Objective听诊器是医疗工作者普遍使用的医疗设备。如果不进行消毒,听诊器可能会在患者和医护人员之间传播病原体。这项研究的目的是确定医护人员使用的听诊器污染程度,调查对听诊器进行消毒的方法,确定各种微生物并评估其作为潜在病原体的作用,并确定70%乙醇作为消毒剂的有效性。 。结果这是在尼泊尔巴拉特普尔的奇旺医学院的微生物学系进行的一项横断面研究。这项研究包括来自不同部门的122名医护人员的听诊器。在总共122个隔膜中,有88个(72.1%)被定殖。只有71(58.1%)的铃铛和152个听筒(66.2%)受到了污染。微球菌和凝固酶阴性葡萄球菌是主要分离的物种。在接触每位患者后清洗的听诊器中,污染最低(11.5%),差异具有统计学意义(P

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号