首页> 外文期刊>Biomedical Optics Express >Depth-resolved model-based reconstruction of attenuation coefficients in optical coherence tomography
【24h】

Depth-resolved model-based reconstruction of attenuation coefficients in optical coherence tomography

机译:光学相干层析成像中基于深度分解模型的衰减系数重建

获取原文
           

摘要

We present a method, based on a single scattering model, to calculate the attenuation coefficient of each pixel in optical coherence tomography (OCT) depth profiles. Numerical simulations were used to determine the model’s response to different depths and attenuation coefficients. Experiments were performed on uniform and layered phantoms with varying attenuation coefficients. They were measured by a 1300 nm OCT system and their attenuation coefficients were evaluated by our proposed method and by fitting the OCT slope as the gold standard. Both methods showed largely consistent results for the uniform phantoms. On the layered phantom, only our proposed method accurately estimated the attenuation coefficients. For all phantoms, the proposed method largely reduced the variability of the estimated attenuation coefficients. The method was illustrated on an in-vivo retinal OCT scan, effectively removing common imaging artifacts such as shadowing. By providing localized, per-pixel attenuation coefficients, this method enables tissue characterization based on attenuation coefficient estimates from OCT data.
机译:我们提出了一种基于单个散射模型的方法,可以计算光学相干断层扫描(OCT)深度剖面中每个像素的衰减系数。数值模拟用于确定模型对不同深度和衰减系数的响应。在具有不同衰减系数的均匀和分层模型上进行了实验。通过1300 nm OCT系统对其进行了测量,并通过我们提出的方法并通过将OCT斜率作为金标准拟合来评估它们的衰减系数。两种方法对于统一体模均显示出基本一致的结果。在分层体模上,只有我们提出的方法可以准确估计衰减系数。对于所有体模,所提出的方法大大降低了估计衰减系数的可变性。该方法在体内视网膜OCT扫描中得到了说明,可有效去除常见的成像伪影,例如阴影。通过提供局部的每个像素的衰减系数,该方法可以根据OCT数据的衰减系数估计值来进行组织表征。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号