Evaluation of reverse osmosis for improving quality of water utilized in hemodialysis devices (case study: Kerman University of Medical Sciences’ hospitals)

机译:进行反渗透评估以改善血液透析设备中使用的水的质量(案例研究:德国克尔曼医科大学医院)

获取原文
获取原文并翻译 | 示例
       

摘要

The presence of some chemical compounds in water used in dialysis devices and the blood circulation system of patients causes many problems, such as acute toxicity, bone and brain diseases. Relevant standards must be observed when valuing these compounds in the water used in a dialysis device. In this study, the rate of total hardness and the concentrations of aluminum, lead, and zinc in dialysis water samples were determined. The concentrations of aluminum, lead, and zinc in the dialysis water at hospitals 1 and 3 and the concentrations of lead and zinc in dialysis water in hospital 2 corresponded with AAMI standards; the concentration of aluminum in the dialysis water at hospitals 1 and 3 also corresponded with EPH standards. Total hardness did not correspond with AAMI standard in any of the three studied hospitals. The promotion of a maintenance program in reverse osmosis devices and the timely replacement of films is important for improving the quality of water for use in dialysis systems.
机译:透析设备和患者血液循环系统中使用的水中存在某些化学化合物会引起许多问题,例如急性毒性,骨骼和脑部疾病。评估透析设备所用水中的这些化合物时,必须遵守相关标准。在这项研究中,确定了总硬度的比率以及透析水样品中铝,铅和锌的浓度。医院1和3的透析水中铝,铅和锌的浓度以及医院2的透析水中铅和锌的浓度符合AAMI标准;医院1和3的透析水中铝的浓度也符合EPH标准。三所研究医院中的任何一家医院的总硬度均不符合AAMI标准。促进反渗透设备的维护计划和及时更换薄膜对于提高用于透析系统的水的质量非常重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号