首页> 外文期刊>Dental materials >Crown pull-off test (crown retention test) to evaluate the bonding effectiveness of luting agents
【24h】

Crown pull-off test (crown retention test) to evaluate the bonding effectiveness of luting agents

机译:牙冠拉脱试验(牙冠保持力试验),以评估胶粘剂的粘合效果

获取原文
获取原文并翻译 | 示例
       

摘要

Objective. The purpose of this review was to assess the influencing factors which affect laboratory tests that evaluate the effectiveness of luting agents on the retention of crowns in prepared dentin and - based on the results of the review - to propose a reasonable experimental setup.rnMaterials and methods. The database MEDLINE was systematically searched for laboratory methods that evaluated the effectiveness of luting agents by pulling off crowns from prepared extracted teeth.rnResults. Eighteen studies were included into the systematic review. The studies varied largely with regard to tooth type (molars, premolars), number of specimens (9-25), stump height (3-6 mm), convergence angle (4.8-33 ), standardization and measurement of preparation surface, seating force (25-200 N), artificial ageing, crosshead speed for tensile force and statistical analysis. The coefficient of variation of the test results varied from 3% to 100%. The most important influencing factors for the crown dislodgement were stump height and convergence angle as well as the luting agent. Panavia and RelyX Unicem generally produced the highest values followed by glass ionomer and zinc phosphate cements. When pooling and normalizing the data, the mean difference between glass ionomer and resin-based materials as well as between glass ionomer and zinc phosphate cements was statistically significant (Wilcoxon, p < 0.05). Seating force, roughness, type of cutting bur and use of a desensitizing agent had all a negligible effect on the test results. Artificial ageing like thermocycling had no influence with glass ionomer cements whereas for resin-based cements thermocycling and prolonged water storage generated similar a failures stress than thermocycling alone. The comparison with clinical results did not reveal conclusive evidence that the results of the laboratory methods completely reflect the results of prospective clinical trials in conjunction with single crowns and fixed dental prostheses. A reasonable experimental setup includes: at least 20 specimens per group, stump height 3 mm, convergence angle 20 , thermocycling of specimens (5000×), avoidance of shearing forces during dislodgement and failure probability statistics (Weibull).rnConclusions. The results of pull-off tests with crowns that are cemented with luting agents varied largely and reflect only partially the results from clinical trials. The most influencing factors (standardized tooth preparation, avoidance of shearing stress during dislodgement of crown) had to be controlled to get meaningful results.
机译:目的。这篇综述的目的是评估影响实验室测试的影响因素,这些指标评估了诱剂对制备的牙本质中牙冠保留效果的有效性,并基于该综述的结果提出了一种合理的实验装置.rn材料和方法。系统搜索了MEDLINE数据库中的实验室方法,这些方法通过从准备好的拔出的牙齿上拔下牙冠来评估诱惑剂的有效性。十八项研究被纳入系统评价。这些研究在牙齿类型(磨牙,前磨牙),标本数量(9-25),树桩高度(3-6 mm),会聚角(4.8-33),准备表面的标准化和测量,座置力方面有很大不同(25-200 N),人工时效,用于拉伸力和统计分析的十字头速度。测试结果的变异系数从3%到100%不等。影响牙冠移位的最重要因素是树桩高度,会聚角以及诱剂。 Panavia和RelyX Unicem通常产生最高值,其次是玻璃离聚物和磷酸锌水泥。当汇总并标准化数据时,玻璃离聚物和树脂基材料之间以及玻璃离聚物和磷酸锌水泥之间的平均差异具有统计学意义(Wilcoxon,p <0.05)。就座力,粗糙度,切削针的类型和使用脱敏剂对测试结果的影响都可以忽略不计。像玻璃纤维热固性水泥这样的人工老化对玻璃离聚物水泥没有影响,而对于树脂基水泥,与单独的热循环相比,热循环和长期储水产生的破坏应力相似。与临床结果的比较并没有得出确凿的证据,表明实验室方法的结果完全反映了单冠和固定义齿结合的前瞻性临床试验的结果。合理的实验设置包括:每组至少20个样本,树桩高度3 mm,会聚角20,样本的热循环(5000×),避免移位时的剪切力和失效概率统计数据(Weibull)。用粘结有粘固剂的牙冠进行拉脱试验的结果差异很大,并且仅部分反映了临床试验的结果。为了获得有意义的结果,必须控制最有影响力的因素(标准化的牙齿准备,避免牙冠移位时的剪切应力)。

著录项

  • 来源
    《Dental materials》 |2010年第3期|193-206|共14页
  • 作者

    S.D. Heintze;

  • 作者单位

    R&D, Iuoclar Vivadent AG, Bendererstrasse 2, FL-9494 Schaan, Liechtenstein;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    crown retention; failure stress; systematic review; luting agent;

    机译:牙冠保持力破坏应力系统评价;诱剂;
  • 入库时间 2022-08-18 03:47:24

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com