首页> 外文期刊>Dental materials >Influence of abutment-to-fixture design on reliability and failure mode of all-ceramic crown systems
【24h】

Influence of abutment-to-fixture design on reliability and failure mode of all-ceramic crown systems

机译:基台到夹具设计对全瓷冠系统可靠性和失效模式的影响

获取原文
获取原文并翻译 | 示例
       

摘要

Objective. Evaluate the effect of implant connection designs on reliability and failure modes of screw-retained all-ceramic crowns. Methods. Central incisor ceramic crowns in zirconia abutments were screwed and torqued down to external hexagon (EH), internal hexagon (IH) and Morse taper (MT) implant systems. Single-load-to-fracture (SLF) test (n = 4 per group) determined three step-stress fatigue profiles with specimens assigned in the ratio of 3:2:1. Fatigue test was performed under water at 10 Hz. Use level probability Weibull curves and reliability for missions of 50,000 cycles at 400 N and 200 N were calculated (90% confidence bounds-CB). Weibull probability distribution (90% CB) was plotted (Weibull modulus vs characteristic strength) for comparison between the groups. Fractographic analyses were conducted under polarized-light microscopy and SEM. Results. Use level Weibull probability calculation indicated that failure was not associated with fatigue in groups EH (β = 0.63), IH (β = 0.97) and MT (β = 0.19). Reliability data for a mission of 50,000 cycles at 400 N revealed significant reliability differences between groups EH (97%), IH (46%) and MT (0.5%) but no significant difference at 200 N between EH (100%) and IH (98%), and IH and MT (89%). Weibull strength distribution (figure) revealed β = 13.1/η = 561.8 for EH, β = 5.8/η = 513.4 for IH and β = 5.3/η = 333.2 for MT. Groups EH and IH exhibited veneer cohesive and adhesive failures. Group IH also presented adhesive failure at zirconia/titanium abutment insert while MT showed fracture at abutment neck. Significance. Although group EH presented higher reliability and characteristic strength followed by IH and MT, all groups withstood reported mean anterior loads.
机译:目的。评估种植体连接设计对螺钉固定的全瓷冠的可靠性和失效模式的影响。方法。氧化锆基台中的中切牙陶瓷冠被拧紧并拧紧到外部六角形(EH),内部六角形(IH)和莫尔斯锥度(MT)植入系统。单载荷断裂试验(每组n = 4)确定了三个阶梯应力疲劳曲线,并按3:2:1的比例分配了试样。疲劳试验在10 Hz的水中进行。计算了使用水平概率威布尔曲线和400 N和200 N下50,000个循环的任务的可靠性(90%置信区间CB)。绘制威布尔概率分布(90%CB)(威布尔模量与特征强度)以进行组间比较。在偏光显微镜和SEM下进行分形分析。结果。使用水平的威布尔概率计算表明,EH(β= 0.63),IH(β= 0.97)和MT(β= 0.19)组的疲劳与疲劳无关。在400 N下进行50,000个循环的任务的可靠性数据显示,EH组(97%),IH(46%)和MT(0.5%)之间的可靠性存在显着差异,而EH(100%)和IH组在200 N时没有显着差异( 98%),以及IH和MT(89%)。威布尔强度分布(图)显示EH的β= 13.1 /η= 561.8,IH的β= 5.8 /η= 513.4和MT的β= 5.3 /η= 333.2。 EH和IH组显示单板内聚和粘合失败。 IH组还显示了氧化锆/钛基台插入件的粘合失败,而MT显示基台颈部断裂。意义。尽管EH组的可靠性和特征强度更高,其次是IH和MT,但所有经受住治疗的组均报告了平均前屈。

著录项

  • 来源
    《Dental materials》 |2014年第4期|408-416|共9页
  • 作者单位

    UNESP, Aracatuba Dental School, Department of Dental Materials and Prosthodontics, Jose Bonifacio, 1193, Aracatuba, SP 16015-050, Brazil,Department of Biomaterials and Biomimetics, New York University, College of Dentistry, New York, NY, USA;

    Department of Prosthodontics and Periodontology, University of Campinas - Piracicaba Dental School, Piracicaba, SP, Brazil;

    Department of Dental Materials and Prosthodontics, Sao Paulo State University - Aracatuba Dental School, Aracatuba, SP, Brazil;

    Department of Restorative Dentistry, Federal University of Minas Gerais - College of Dentistry, Belo Horizonte, MG, Brazil;

    Department of Dental Materials, King's College - London Dental Institute, London, UK;

    Department of Dental Materials and Prosthodontics, Sao Paulo State University - Aracatuba Dental School, Aracatuba, SP, Brazil;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Reliability; Dental implants; Zirconia; Implant-supported prosthesis;

    机译:可靠性;牙种植体;氧化锆植入物支持的假体;
  • 入库时间 2022-08-18 03:46:56

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号