STUDY ON A PRECISE ON-SPOT CALIBRATION METHOD FOR DISTRIBUTED STEREO REFERENCE POSITION-SETTING

机译:分布式立体参考位置设置的精确点校准方法研究

获取原文
获取原文并翻译 | 示例
           

摘要

A high precision method used for on-spot calibration of distributed stereo reference position setting is presented. The high measuring accuracy in stereo reference calibration is derived from using a high precision water level instrument and an accurate height vernier caliper. It settles the problem of reference calibration effectively and accurately, without using large coordinate measuring machines (CMM). It is more adaptable and precise than traditional calibration methods applying theodolites or autocollimators. The error sources of this method are analyzed in detail and several methods are developed to eliminate the calibration error.An optimizing swallowtail-like anchor target is developed. Experiments show that the calibration accuracy can be limited within 0.06 mm in the range of 3~5 m and 0.03 mm with optimizing anchor target. This method can be widely used in on-spot calibration.
机译:提出了一种用于现场校准分布式立体声参考位置设置的高精度方法。立体参考校准的高测量精度源自使用高精度水位仪和精确的高度游标卡尺。它无需使用大型坐标测量机(CMM),即可有效,准确地解决参考校准的问题。与使用经纬仪或自动准直仪的传统校准方法相比,它更具适应性和准确性。详细分析了该方法的误差源,并提出了几种消除校准误差的方法。实验表明,通过优化锚定目标,可以将标定精度限制在3〜5 m至0.03 mm范围内的0.06 mm以内。该方法可广泛用于现场校准。