首页> 外文期刊>OIML bulletin >Method for calibrating line standards with marks, with nominal length up to 1000 mm: Uncertainty budget
【24h】

Method for calibrating line standards with marks, with nominal length up to 1000 mm: Uncertainty budget

机译:标定最大长度为1000 mm的带有标记的线标准的校准方法:不确定度预算

获取原文
获取原文并翻译 | 示例
           

摘要

This paper describes a modern method used for the calibration of line standards with marks using a longitudinal comparator with an He-Ne frequency-stabilized laser source, and the subsequent estimation of the errors which occur during the measurement of line standards with such a longitudinal comparator. The author's contribution consists of: 1. the analyses of the sources of error which occur during the measurement due to the line standard, longitudinal comparator, electronic equipment and temperature; 2. the estimation of the measurement uncertainty budget according to the specifications of the GUM; and 3. finding a better longitudinal comparator construction solution to increase the performance of the measurement of line standards, consisting in replacing the optical microscope with a modern viewing model equipped with a CCD camera and custom software for capturing and analyzing the image.
机译:本文介绍了一种现代方法,该方法用于使用带有He-Ne频率稳定激光源的纵向比较器对带有标记的线标准进行校准,并随后估算使用这种纵向比较器测量线标准时出现的误差。作者的贡献包括:1.分析测量中由于线路标准,纵向比较器,电子设备和温度而引起的误差源; 2.根据GUM的规范估算测量不确定度预算; 3.寻找更好的纵向比较器构造解决方案以提高线标准的测量性能,包括用配备CCD摄像头和用于捕获和分析图像的定制软件的现代观察模型代替光学显微镜。