We have developed a novel noncontact respiration monitoring that utilized an ultrasonic phase-shift reconstruction method for measuring the normal vibration of chest-wall. In this paper, a simple but efficient method for vibration measurement is adopted for the sleep apnea syndrome. The instrument consists of a pair of 40 kHz ultrasonic transducers which are used in a transmit-receive mode to get the motion of a point on a vibration object. The relative motion of the object modulates the phase angle between transmitted and received ultrasound signals. We have also developed an ingenious method to reconstruct the relative motion of an object from the acquired data of the phase angle changes. The measurement accuracy of the system in the reported experiments is within +/- 0.07 mm. The main advantages of this vibration measurement system are high resolution, low cost, and noncontact measurement.
机译:我们开发了一种新颖的非接触式呼吸监测,该监测利用超声相移重建方法来测量胸壁的正常振动。本文针对睡眠呼吸暂停综合症采用了一种简单而有效的振动测量方法。该仪器由一对40 kHz超声换能器组成,该超声换能器用于收发模式,以获取振动物体上一点的运动。物体的相对运动调节发射和接收的超声信号之间的相位角。我们还开发了一种巧妙的方法,可以从获取的相角变化数据中重建对象的相对运动。在所报告的实验中,系统的测量精度在+/- 0.07 mm之内。这种振动测量系统的主要优点是高分辨率,低成本和非接触式测量。