Use of impedance ratio for the continuous measurement of stroke volume of a valveless pouch used as a cardiac-assist device

机译:阻抗比用于连续测量用作心脏辅助装置的无瓣膜袋的搏动量

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A technique is described for measuring the volume of a valveless compressible plastic pouch and its volume change when used as a cardiac-assist device. The method employs measuring the pouch impedance at high frequency with sleeve electrodes at both ends of the pouch. The use of an adequately high frequency eliminates the electrode impedance, and the impedance measured is that of the resistance of the electrolyte in the pouch. By equating the compressible pouch to two truncated cones with their bases adjacent, an equation is derived that relates pouch impedance to volume. It is shown that by plotting the stroke volume ejected ( Delta V) versus the ratio of systolic (R/sub S/) to diastolic (R/sub d/) impedance, the resulting relationship is independent of the resistivity of the fluid in the pouch. Validation tests were made with a 100 mL pouch filled with solutions having resistivities of 60, 102, 145, and 192 Omega -cm. The method permits calibration of the volume of a valveless pouch used as a circulatory-assist device regardless of changes in packed-cell volume.
机译:描述了一种用于测量无阀可压缩塑料袋的体积及其在用作心脏辅助装置时的体积变化的技术。该方法采用在袋的两端使用套筒电极以高频测量袋的阻抗。使用足够高的频率消除了电极阻抗,并且所测量的阻抗是袋中电解质电阻的阻抗。通过将可压缩袋等同于两个圆锥台,使它们的底部相邻,可以得出将袋阻抗与体积相关的方程。结果表明,通过绘制排出的搏动量(Delta V)与收缩压(R / sub S /)与舒张压(R / sub d /)的比率的关系图,得出的关系与流体在导管中的电阻率无关。袋。用装满电阻率为60、102、145和192Ω-cm的溶液的100 mL袋进行验证测试。该方法允许校准用作循环辅助装置的无阀袋的体积,而与填充细胞体积的变化无关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号