The sensitivity of an impedance imaging system to small cylindrical inhomogeneities in two-dimensional (2-D) and three-dimensional (3-D) saline tanks was studied for different height electrodes and different height targets. Experimental results were compared with analytical models. Inhomogeneities in the 3-D tank having limited vertical extent were detected by electrodes of vertical size comparable to that of the inhomogeneity. Taller electrodes had increased sensitivity to short targets to only a limited extent. When the electrode height was more than twice that of the target, sensitivity decreased or remained the same with further increases in electrode height. The system was less sensitive to inhomogeneities in the 3-D tank than to those in the 2-D tank. The distinguishability of conductors was greater than that of insulators in the 3-D tank, and the opposite was true in the 2-D tank, consistent with an analytical result.
机译:针对不同高度的电极和不同的高度目标,研究了阻抗成像系统对二维(2-D)和三维(3-D)盐水罐中小的圆柱形不均匀性的敏感性。实验结果与分析模型进行了比较。通过垂直尺寸可与不均匀性相当的电极检测到垂直范围有限的3-D储罐中的不均匀性。较高的电极仅在有限的程度上提高了对短靶的敏感性。当电极高度超过目标高度的两倍时,灵敏度降低或保持不变,电极高度进一步增加。该系统对3D储罐中的不均匀性的敏感度要比对2D储罐中的不均匀性敏感。与分析结果一致,在3-D槽中,导体的可分辨性大于绝缘子,在2-D槽中则相反。