首页> 外文期刊>IEEE Transactions on Biomedical Engineering >Catheter-Induced Errors in Pressure Measurements in Vessels: An In-Vitro and Numerical Study
【24h】

Catheter-Induced Errors in Pressure Measurements in Vessels: An In-Vitro and Numerical Study

机译:导管内压力测量中导管引起的误差:体外和数值研究

获取原文
获取原文并翻译 | 示例

摘要

Accurate measurement of blood pressure is important because it is a biomarker for cardiovascular disease. Diagnostic catheterization is routinely used for pressure acquisition in vessels despite being subject to significant measurement errors. To investigate these errors, this study compares pressure measurement using two different techniques in vitro and numerical simulations. Pressure was acquired in a pulsatile flow phantom using a 6F fluid-filled catheter and a 0.014′′ pressure wire, which is considered the current gold standard. Numerical simulations of the experimental set-up with and without a catheter were also performed. Despite the low catheter-to-vessel radius ratio, the catheter traces showed a 24% peak systolic pressure overestimation compared to the wire. The numerical models replicated this difference and indicated the cause for overestimation was the increased flow resistance due to the presence of the catheter. Further, the higher frequency pressure oscillations observed in the wire and numerical data were absent in the catheter, resulting in an overestimation of the pulse wave velocity with the latter modality. These results show that catheter geometry produces significant measurement bias in both the peak pressure and the waveform shape even with radius ratios considered acceptable in clinical practice. The wire allows for more accurate pressure quantification, in agreement with the numerical model without a catheter.
机译:准确测量血压很重要,因为它是心血管疾病的生物标记。尽管存在明显的测量误差,诊断性导管插入术仍常规用于血管压力采集。为了调查这些错误,本研究比较了使用两种不同的体外和数值模拟技术进行的压力测量。使用6F充液导管和0.014''压力线在脉动流动体模中获取压力,这被认为是当前的金标准。还进行了带有和不带有导管的实验装置的数值模拟。尽管导管与血管的半径比很低,但导管导线显示的收缩压峰值仍比导线高24%。数值模型复制了这种差异,并指出了高估的原因是由于导管的存在导致流动阻力增加。此外,在导管中不存在在导线中观察到的较高频率的压力振荡和数值数据,从而导致在后者的情况下高估了脉搏波速度。这些结果表明,即使在临床实践中认为可接受的半径比,导管的几何形状也会在峰值压力和波形形状上产生明显的测量偏差。与没有导管的数值模型相一致,金属丝允许更精确的压力定量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号