首页> 外文期刊>Biomedical Chromatography >Determination of ibudilast in human serum by high-performance liquid chromatography for pharmacokinetic study
【24h】

Determination of ibudilast in human serum by high-performance liquid chromatography for pharmacokinetic study

机译:高效液相色谱法测定人血清中异丁司特的药代动力学研究

获取原文
           

摘要

A simple, accurate, precise and cost effective reversed-phase HPLC method was developed to determine the concentration of ibudilast in human serum. Ibudilast and an internal standard, butyl 4-hydroxybenzoate, were extracted by liquid–liquid extraction with methyl tert-butyl ether. HPLC analysis was carried out under the following conditions: a Luna C18(2) 5 μm column, a mobile phase of acetonitrile–0.02% phosphoric acid (50 : 50, v/v, adjusted to pH 6.0 with triethylamine) and a UV detector at 319 nm. The chromatograms showed good resolution and sensitivity as well as no interference from the human serum. The calibration curves were linear over the concentration range, 1–100 ng/mL, for serum with correlation coefficients 0.999. The intra- and inter-day assay precision as well as the accuracy fulfilled the international requirements. The mean absolute recovery for human serum was 101.7 ± 6.1%. The lower limit of quantitation in human serum was 1 ng/mL, which is sensitive enough for pharmacokinetic studies. Stability studies revealed that ibudilast in human serum was stable during storage as well as during the assay procedure. This method was applied successfully to an examination of the pharmacokinetics of ibudilast in human subjects following a single oral dose of an ibudilast (10 mg) capsule. Copyright © 2009 John Wiley & Sons, Ltd.
机译:开发了一种简单,准确,精确且具有成本效益的反相HPLC方法,用于测定人血清中异丁司特的浓度。用甲基叔丁基醚进行液-液萃取,提取出Ibudilast和内标4-羟基苯甲酸丁酯。 HPLC分析是在以下条件下进行的:Luna C 18 (2)5μm色谱柱,乙腈-0.02%磷酸(50 phase:50,v / v,调节至pH值)的流动相6.0(含三乙胺)和319 nm的UV检测器。色谱图显示出良好的分离度和灵敏度,并且不受人血清的干扰。对于相关系数> 0.999的血清,校准曲线在浓度范围1–100 ng / mL内呈线性。日内和日间测定的准确性以及准确性均达到国际要求。人血清的平均绝对回收率为101.7±6.1%。人血清中的定量下限为1μng/ mL,对药代动力学研究足够敏感。稳定性研究表明,人血清中的ibudilast在储存以及测定过程中均稳定。该方法已成功应用于单次口服ibudilast(10 mg)胶囊后在人受试者中ibudilast的药代动力学检查。版权所有©2009 John Wiley&Sons,Ltd.

著录项

  • 来源
    《Biomedical Chromatography》 |2010年第3期|p.324-328|共5页
  • 作者单位

    College of Pharmacy and Institute of Bioequivalence and Bridging Study, Chonnam National University, 300 Yongbong-Dong, Buk-Gu, Gwangju 500-757, Korea;

    Department of Pharmacological Research, NITR, KFDA, Seoul 122-704, Korea;

    College of Pharmacy and Institute of Bioequivalence and Bridging Study, Chonnam National University, 300 Yongbong-Dong, Buk-Gu, Gwangju 500-757, Korea;

  • 收录信息
  • 原文格式 PDF
  • 正文语种
  • 中图分类
  • 关键词

    ibudilast; human serum; HPLC; pharmacokinetics;

    机译:异丁司特;人血清;HPLC;药代动力学;