A comparison between EPSON V700 and EPSON V800 scanners for film dosimetry

机译:EPSON V700和EPSON V800扫描仪对胶片剂量测定的比较

获取原文
获取原文并翻译 | 示例
           

摘要

Radiochromic film is a good dosimeter choice for patient QA for complex treatment techniques (IMRT, VMAT, SABR, SBRT) because of its near tissue equivalency, very high spatial resolution and established method of use. Commercial scanners are usually used for film dosimetry, among which EPSON scanners are the most common. NCCI have used an EPSON V700 scanner, but recently acquired a new model EPSON V800 scanner. The purpose of this work was to evaluate any differences between these two scanners to consider whether they can be used interchangeably or not. Different aspects of film dosimetry, e.g. lateral response artefact (LRA) effect, orientation effect, scanner response etc., were compared. EBT3 films were irradiated with 40 x 40 cm(2) field size 6 MV beams and scanned in both the scanners. The scanned images were read in ImageJ V1.49 software. The data obtained was then copied in MS Excel to compare the scanners. The V800 scanner causes more polarisation, which results in more LRA effect than for the V700 scanner. The responses of the scanners in all three colour channels are not the same for the same film and irradiation. The V800 scanner shows an increase of response of up to 1.6% compared to 3.7% increase in the V700 scanner after scanning a piece of irradiated film 20 times. The scanners cannot be used interchangeably. The correction factors for LRA effect and the calibration curves are different. Further characterisation, evaluation and commissioning is required before clinical use.
机译:放射刻度薄膜是患者QA的良好剂量计,适用于复杂的处理技术(IMRT,VMAT,SABR,SBRT),因为其近乎组织等同,非常高的空间分辨率和建立的使用方法。商业扫描仪通常用于薄膜剂量测定法,其中EPSON扫描仪是最常见的。 NCCI使用了EPSON V700扫描仪,但最近收购了一个新型EPSON V800扫描仪。这项工作的目的是评估这两个扫描仪之间的任何差异,以考虑它们是否可以互换使用。薄膜剂量法的不同方面,例如,比较横向响应人工制品(LRRA)效应,取向效果,扫描仪响应等。 EBT3薄膜用40×40cm(2)场尺寸6 mV梁照射并在扫描仪中扫描。扫描图像在imagej v1.49软件中读取。然后在MS Excel中复制所获得的数据以比较扫描仪。 V800扫描仪会导致更多的极化,从而导致比V700扫描仪更高的LRA效果。对于相同的胶片和辐射,所有三种颜色通道中扫描仪的响应不相同。 V800扫描仪的增加响应高达1.6%,而V700扫描仪扫描一块辐照膜20次后,频率增加3.7%。扫描仪不能互换使用。 LRA效应和校准曲线的校正因子是不同的。在临床用途之前需要进一步表征,评估和调试。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号