Equipotential earthing study: equipotential wiring is unnecessary

机译:等电位接地研究:不需要等电位接线

获取原文
获取原文并翻译 | 示例
           

摘要

This study was undertaken to determine the ongoing need for Equipotential (EP) earthing as required in AS/NZS 3003, section 4.4.2. A total of 2279 earth points were tested in four clinical institutions in Body Protected areas (as defined in AS/NZS 3003). Each earth point was tested for potentials existing in the earthing in order to assess whether any potentials exceeded the 100 mV requirement of AS/NZS 3003 for EP earthing. The voltage measured was 1.21 mV +/-3.59 mV, (mean +/- standard deviation (SD)), with a of 99 % confidence interval between 1.0 mV and 1.4 mV. This study suggests that the requirements for EP earthing in section 4.4.2 of AS/NZS 3003 are unnecessary. It is further suggested that Standards Australia review these requirements with the view of withdrawing them from the Standard.
机译:进行这项研究是为了确定AS / NZS 3003第4.4.2节所要求的等电位(EP)接地的持续需求。在人体保护区的4个临床机构(按照AS / NZS 3003的定义)中总共测试了2279个地球点。测试每个接地点的接地电位,以评估是否有任何电位超出AS / NZS 3003的EP接地的100 mV要求。测得的电压为1.21 mV +/- 3.59 mV(平均+/-标准偏差(SD)),在1.0 mV和1.4 mV之间具有99%的置信区间。这项研究表明,AS / NZS 3003的4.4.2节中有关EP接地的要求是不必要的。进一步建议澳大利亚标准协会审查这些要求,以将其从标准中撤回。