Sensitivity of an Elekta iView GT a-Si EPID model to delivery errors for pre-treatment verification of IMRT fields

机译:Elekta iView GT a-Si EPID模型对传递错误的敏感性,用于IMRT场的预处理验证

获取原文
获取原文并翻译 | 示例

摘要

A Monte Carlo model of an Elekta iViewGT amorphous silicon electronic portal imaging device (a-Si EPID) has been validated for pre-treatment verification of clinical IMRT treatment plans. The simulations involved the use of the BEAMnrc and DOSXYZnrc Monte Carlo codes to predict the response of the iViewGT a-Si EPID model. The predicted EPID images were compared to the measured images obtained from the experiment. The measured EPID images were obtained by delivering a photon beam from an Elekta Synergy linac to the Elekta iViewGT a-Si EPID. The a-Si EPID was used with no additional build-up material. Frame averaged EPID images were acquired and processed using in-house software. The agreement between the predicted and measured images was analyzed using the gamma analysis technique with acceptance criteria of 3 %/3 mm. The results show that the predicted EPID images for four clinical IMRT treatment plans have a good agreement with the measured EPID signal. Three prostate IMRT plans were found to have an average gamma pass rate of more than 95.0 % and a spinal IMRT plan has the average gamma pass rate of 94.3 %. During the period of performing this work a routine MLC calibration was performed and one of the IMRT treatments re-measured with the EPID. A change in the gamma pass rate for one field was observed. This was the motivation for a series of experiments to investigate the sensitivity of the method by introducing delivery errors, MLC position and dosimetric overshoot, into the simulated EPID images. The method was found to be sensitive to 1 mm leaf position errors and 10 % overshoot errors.
机译:Elekta iViewGT非晶硅电子门成像设备(a-Si EPID)的蒙特卡洛模型已经过验证,可用于临床IMRT治疗计划的治疗前验证。仿真涉及使用BEAMnrc和DOSXYZnrc蒙特卡洛代码来预测iViewGT a-Si EPID模型的响应。将预测的EPID图像与从实验中获得的测量图像进行比较。通过将光子束从Elekta Synergy直线加速器传递到Elekta iViewGT a-Si EPID可获得测得的EPID图像。使用a-Si EPID,无需额外的堆积材料。使用内部软件获取并处理了帧平均的EPID图像。使用伽玛分析技术以3%/ 3 mm的接受标准分析了预测图像和测量图像之间的一致性。结果表明,针对四种临床IMRT治疗计划的预测EPID图像与测得的EPID信号具有良好的一致性。发现三个前列腺IMRT计划的平均伽玛通过率超过95.0%,而脊柱IMRT计划的平均伽玛通过率达到94.3%。在执行此工作期间,将执行常规的MLC校准,并使用EPID重新测量IMRT处理之一。观察到一个场的伽马通过率的变化。这是一系列实验的动机,这些实验通过将传递误差,MLC位置和剂量超调引入模拟的EPID图像中来研究该方法的敏感性。发现该方法对1 mm叶片位置误差和10%过冲误差敏感。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号