Radiation dose measurements of an on-board imager X-ray unit using optically-stimulated luminescence dosimeters

机译:使用光激发发光剂量计测量机载成像仪X射线单元的辐射剂量

获取原文
获取原文并翻译 | 示例
           

摘要

Cone beam computed tomography (CBCT) is now widely used to image radiotherapy patients prior to treatment for the purpose of accurate patient setup. However each CBCT image delivered to a patient increases the total radiation dose that they receive. The measurement of the dose delivered from the CBCT images is not readily performed in the clinic. In this study, we have used commercially available optically stimulated luminescence (OSLD) dosimeters to measure the dose delivered by the Varian OBI on a radiotherapy linear accelerator. Calibration of the OSLDs was achieved by using a therapeutic X-ray unit. The dose delivered by a head CBCT scan was found to be 3.2 +/- A 0.3 mGy which is similar in magnitude to the dose of a head computed tomography (CT) scan. The results of this study suggest that the radiation hazard associated with CBCT is of a similar nature to that of conventional CT scans. We have also demonstrated that the OSLDs are suitable for these low X-ray dose measurements.
机译:锥形束计算机断层扫描(CBCT)现在被广泛用于对放射治疗前的患者进行成像,以实现准确的患者设置。但是,传递给患者的每个CBCT图像都会增加他们接收的总辐射剂量。从CBCT图像递送的剂量的测量在临床中不容易进行。在这项研究中,我们使用了市售的光激发发光(OSLD)剂量计来测量Varian OBI在放射治疗线性加速器上的剂量。 OSLD的校准是通过使用X射线治疗仪实现的。发现头部CBCT扫描的剂量为3.2 +/- A 0.3 mGy,其大小与头部计算机断层扫描(CT)扫描的剂量相似。这项研究的结果表明,与CBCT相关的辐射危害与常规CT扫描具有相似的性质。我们还证明了OSLD适用于这些低X射线剂量测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号