A comparison between direct TMR measurements and TMRs calculated from PDDs using BJR Supplement 25 data for flattened and unflattened photon beams

机译:直接TMR测量与使用BJR补充25数据针对平坦和未平坦光子束从PDD计算得出的TMR之间的比较

获取原文
获取原文并翻译 | 示例
           

摘要

This study assessed the validity of the conversion from percentage depth dose (PDD) to tissue maximum ratio (TMR) using BJR Supplement 25 data for flattened and flattening filter free (FFF) beams. PDD and TMR scans for a variety of field sizes were measured in water using a Sun Nuclear Corporation 3D SCANNER (TM) on a Varian TrueBeam linear accelerator in 6 MV, 10 MV and 6 MV FFF beams. The BJR Supplement 25 data was used to convert the measured PDDs to TMRs and these were compared with the directly measured TMR data. The TMR plots calculated from PDD were within 1 % for the 10 MV and 6 MV flattened beams, for field sizes 3 cm x 3 cm to 40 cm x 40 cm inclusive, at depths measured beyond the depth of maximum dose. The disagreement between the measured and calculated TMR plots for the 6 MV FFF beam increased with depth and field size to a maximum of 1.7 % for a 40 cm x 40 cm field. The results found in this study indicate that the BJR Supplement 25 data should not be used for field sizes larger than 20 cm x 20 cm at depths greater than 15 cm for the 6 MV FFF beam. It is advised that PDD to TMR conversion for FFF beams should be done with phantom scatter ratios appropriate to FFF beams, or the TMR should be directly measured if required.
机译:这项研究使用BJR补充25数据对平坦和平坦的无滤光器(FFF)光束评估了从百分比深度剂量(PDD)转换为组织最大比率(TMR)的有效性。在6 MV,10 MV和6 MV FFF光束中,使用Sun Nuclear Corporation 3D SCANNER(TM)在Varian TrueBeam线性加速器上在水中测量了各种场大小的PDD和TMR扫描。 BJR Supplement 25数据用于将测得的PDD转换为TMR,并将其与直接测得的TMR数据进行比较。对于10 MV和6 MV平坦光束,从PDD计算出的TMR图在1 cm以内,对于3 cm x 3 cm至40 cm x 40 cm(包括两端)的场大小,其测量深度超过最大剂量的深度。对于6 MV FFF光束,测得的TMR图和计算得出的TMR图之间的差异随着深度和场大小的增加而增加,对于40 cm x 40 cm场,最大值为1.7%。在这项研究中发现的结果表明,对于6 MV FFF光束,当深度大于15 cm时,对于大于20 cm x 20 cm的场尺寸,不应使用BJR补充25数据。建议FFF光束的PDD到TMR的转换应采用适合FFF光束的幻影散射比,或者如果需要,应直接测量TMR。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号