首页> 外文期刊>Australasian physical & engineering sciences in medicine >Evaluating the effect of various intracavitary applicators on dosimetric parameters of Ir-192, Cs-137, and Co-60 sources
【24h】

Evaluating the effect of various intracavitary applicators on dosimetric parameters of Ir-192, Cs-137, and Co-60 sources

机译:评估各种腔内涂药器对Ir-192,Cs-137和Co-60放射源剂量参数的影响

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The purpose of this research is to study the effect of various applicator compositions on dosimetric parameters and dose distribution of Ir-192, Cs-137, and Co-60 sources, using Monte Carlo simulation techniques. To study the effect of applicators on source dosimetry, the dose rate constant, and radial dose function and isodose curves for the above noted sources were calculated in the presence and absence of plastic, titanium, and a stainless steel applicators. The effects of the applicators on the dosimetric parameters and isodose curves of these sources were dependent of the source type and materials of the applicator. The Ir-192 source with the stainless steel applicator has the maximum difference of dose rate (4.2 %) relative to the without applicator case. The Co-60 source with plastic applicator has the minimum dose variation. Moreover, this effect is higher for lower energy sources. Ignoring the effect of applicator composition and geometry on dose distribution may cause discrepancies in treatment planning. Plastic applicators have the least radiation attenuation compared to the other applicators, therefore, they are recommended for use in brachytherapy. A table of correction factors has been introduced for different sources and applicators with different materials for the clinical applications.
机译:这项研究的目的是使用蒙特卡罗模拟技术研究各种涂药器组成对剂量参数和Ir-192,Cs-137和Co-60离子源剂量分布的影响。为了研究施药器对放射源剂量测定的影响,在有无塑料施药器,钛施药器和不锈钢施药器的情况下,计算了上述放射源的剂量率常数,径向剂量函数和等剂量曲线。施加器对这些源的剂量参数和等剂量曲线的影响取决于施加器的源类型和材料。相对于不带涂抹器的情况,带有不锈钢涂抹器的Ir-192放射源的剂量率差异最大(4.2%)。带有塑料喷头的Co-60辐射源具有最小的剂量变化。此外,对于较低的能源,这种效果更高。忽略施药器组成和几何形状对剂量分布的影响可能会导致治疗计划中的差异。与其他涂抹器相比,塑料涂抹器的辐射衰减最小,因此,建议将其用于近距离放射治疗。已针对不同来源和使用不同材料的涂抹器为临床应用引入了校正系数表。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号