首页> 外文期刊>IEEE Transactions on Applied Superconductivity >Multi-Channel SQUID System for MEG and Ultra-Low-Field MRI
【24h】

Multi-Channel SQUID System for MEG and Ultra-Low-Field MRI

机译:用于MEG和超低场MRI的多通道SQUID系统

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A seven-channel system capable of performing both magnetoencephalography (MEG) and ultra-low-field magnetic resonance imaging (ULF MRI) is described. The system consists of seven second-order SQUID gradiometers with 37 mm diameter and 60 mm baseline, having magnetic field resolution of 1.2–2.8 ${rm fT}/surd{rm Hz}$ . It also includes four sets of coils for 2-D Fourier imaging with pre-polarization. The system''''s MEG performance was demonstrated by measurements of auditory evoked response. The system was also used to obtain a multi-channel 2-D image of a whole human hand at the measurement field of 46 microtesla with 3 by 3 mm resolution.
机译:描述了一种能够执行脑磁图(MEG)和超低场磁共振成像(ULF MRI)的七通道系统。该系统由七个直径为37 mm,基线为60 mm的二阶SQUID梯度仪组成,其磁场分辨率为1.2–2.8 $ {rm fT} / surd {rm Hz} $。它还包括四组线圈,用于带有预极化的二维傅立叶成像。该系统的MEG性能通过听觉诱发反应的测量来证明。该系统还用于在46微特斯拉的测量场上以3 x 3毫米的分辨率获得整个人手的多通道2D图像。